Experimental Evaluation of the Optimal Suture Pattern With a Flexible Endoscopic Suturing System. (June 2017)
- Record Type:
- Journal Article
- Title:
- Experimental Evaluation of the Optimal Suture Pattern With a Flexible Endoscopic Suturing System. (June 2017)
- Main Title:
- Experimental Evaluation of the Optimal Suture Pattern With a Flexible Endoscopic Suturing System
- Authors:
- Halvax, Peter
Diana, Michele
Nagao, Yoshihiro
Marescaux, Jacques
Swanström, Lee - Abstract:
- Background . The ability to perform reliable, secure endoluminal closure of the gastrointestinal tract wall, is a prerequisite to support the progress of the emerging field of endoluminal surgery. Along with advanced clipping systems, flexible endoscopic suturing devices are commercially available. Current systems can replicate traditional surgical suturing patterns in the endoluminal environment. The aim of this study was to evaluate the optimal endoluminal suturing technique using a flexible endoscopic suturing device. Materials and Methods . Procedures were performed on bench-top simulators containing 20 explanted porcine stomachs. A standardized 3-cm full-thickness incision was created on the anterior wall of each stomach using monopolar cautery. The gastrotomy was closed endoscopically using an over-the-scope suturing device (OverStitch, Apollo Endosurgery; Austin, TX). Three different techniques were used: single stitches, figure-of-8 pattern, and running suture. Material consumption and operation time were recorded and bursting pressure measurement of the closure was performed. Results . No statistically significant differences were identified in suturing time. Suturing time (minutes) was slightly shorter with the figure-of-8 technique (41.14 ± 4.6) versus interrupted (45.75 ± 1.1) versus continuous (51.44 ± 10.0), but the difference was not statistically significant. The number of sutures required was greater in the interrupted group. No significant difference wasBackground . The ability to perform reliable, secure endoluminal closure of the gastrointestinal tract wall, is a prerequisite to support the progress of the emerging field of endoluminal surgery. Along with advanced clipping systems, flexible endoscopic suturing devices are commercially available. Current systems can replicate traditional surgical suturing patterns in the endoluminal environment. The aim of this study was to evaluate the optimal endoluminal suturing technique using a flexible endoscopic suturing device. Materials and Methods . Procedures were performed on bench-top simulators containing 20 explanted porcine stomachs. A standardized 3-cm full-thickness incision was created on the anterior wall of each stomach using monopolar cautery. The gastrotomy was closed endoscopically using an over-the-scope suturing device (OverStitch, Apollo Endosurgery; Austin, TX). Three different techniques were used: single stitches, figure-of-8 pattern, and running suture. Material consumption and operation time were recorded and bursting pressure measurement of the closure was performed. Results . No statistically significant differences were identified in suturing time. Suturing time (minutes) was slightly shorter with the figure-of-8 technique (41.14 ± 4.6) versus interrupted (45.75 ± 1.1) versus continuous (51.44 ± 10.0), but the difference was not statistically significant. The number of sutures required was greater in the interrupted group. No significant difference was found in the burst pressure (mm Hg): figure-of-8 (45.85 ± 26.2) versus interrupted (30.5 ± 22.89) versus continuous (32.0 ± 26.5). In the figure-of-8 group, 85.5% of cases were leakproof above 30 mm Hg, while in the other groups only 50% of cases were so. Conclusion . A figure-of-8 suturing pattern seems to be the preferable suturing technique with the endoscopic suturing device. … (more)
- Is Part Of:
- Surgical innovation. Volume 24:Number 3(2017:Jun.)
- Journal:
- Surgical innovation
- Issue:
- Volume 24:Number 3(2017:Jun.)
- Issue Display:
- Volume 24, Issue 3 (2017)
- Year:
- 2017
- Volume:
- 24
- Issue:
- 3
- Issue Sort Value:
- 2017-0024-0003-0000
- Page Start:
- 201
- Page End:
- 204
- Publication Date:
- 2017-06
- Subjects:
- endoluminal suturing device -- suturing techniques -- OverStitch -- endoscopy -- perforation -- suture -- closure
Surgery, Operative -- Periodicals
Endoscopic surgery -- Periodicals
Laparoscopic surgery -- Periodicals
Surgical Procedures, Operative -- Periodicals
Surgical Procedures, Minimally Invasive -- Periodicals
Diffusion of Innovation -- Periodicals
Chirurgie opératoire -- Périodiques
Chirurgie endoscopique -- Périodiques
Chirurgie laparoscopique -- Périodiques
617.91 - Journal URLs:
- http://journals.sagepub.com/home/sri ↗
http://sri.sagepub.com/ ↗
http://www.sagepub.com/journalsProdDesc.nav?prodId=Journal201793 ↗
http://www.sagepublications.com/ ↗ - DOI:
- 10.1177/1553350617697184 ↗
- Languages:
- English
- ISSNs:
- 1553-3506
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7590.xml