Quality-controlled dose-reduction of pelvic X-ray examinations in infants with hip dysplasia. Issue 1 (January 2016)
- Record Type:
- Journal Article
- Title:
- Quality-controlled dose-reduction of pelvic X-ray examinations in infants with hip dysplasia. Issue 1 (January 2016)
- Main Title:
- Quality-controlled dose-reduction of pelvic X-ray examinations in infants with hip dysplasia
- Authors:
- Kloth, Jost Karsten
Neumann, Regina
von Stillfried, Eva
Stiller, Wolfram
Burkholder, Iris
Kauczor, Hans-Ulrich
Ewerbeck, Volker
Weber, Marc-André - Abstract:
- Highlights: Pelvic radiographs of infants with hip dysplasia with speed class 400 and 800 were compared. Radiographs with reduced dose (800) were not inferior to those with standard dose (400). We recommend speed class 800 as new reference parameter in these examinations. Abstract: Background: Digital plain radiography (DR) examinations of the pelvis are frequently performed in infants with hip dysplasia. Objective: The purpose was to reduce the radiation dose and to determine objective quality control criteria to ensure accurate assessment. This seems feasible because of higher quantum efficiency of DR and easy assessable anatomical structures for most orthopaedic measurements. Materials and methods: Institutional review board approval was obtained. In this prospective randomized study, 264 patients underwent X-ray examination of the pelvis with standard and reduced dose. The evaluation of the plain-radiographs was conducted using the following criteria: acetabular and center edge angle, closing of the epiphyseal plates and maturation of the femoral head. Two radiologists evaluated these criteria using a score ranging from 1 (definitely assessable) to 4 (not assessable). If a single criterion had been evaluated with a score of 3 or more points or more than 2 criteria with 2 points, the radiograph was scored as "not assessable". The statistical analysis was conducted as non-inferiority-trial. Results: Five (1.9%) examined X-rays were scored as not assessable. There was noHighlights: Pelvic radiographs of infants with hip dysplasia with speed class 400 and 800 were compared. Radiographs with reduced dose (800) were not inferior to those with standard dose (400). We recommend speed class 800 as new reference parameter in these examinations. Abstract: Background: Digital plain radiography (DR) examinations of the pelvis are frequently performed in infants with hip dysplasia. Objective: The purpose was to reduce the radiation dose and to determine objective quality control criteria to ensure accurate assessment. This seems feasible because of higher quantum efficiency of DR and easy assessable anatomical structures for most orthopaedic measurements. Materials and methods: Institutional review board approval was obtained. In this prospective randomized study, 264 patients underwent X-ray examination of the pelvis with standard and reduced dose. The evaluation of the plain-radiographs was conducted using the following criteria: acetabular and center edge angle, closing of the epiphyseal plates and maturation of the femoral head. Two radiologists evaluated these criteria using a score ranging from 1 (definitely assessable) to 4 (not assessable). If a single criterion had been evaluated with a score of 3 or more points or more than 2 criteria with 2 points, the radiograph was scored as "not assessable". The statistical analysis was conducted as non-inferiority-trial. Results: Five (1.9%) examined X-rays were scored as not assessable. There was no statistical inferiority between the examinations with standard (4.57 μSv) or reduced dose (3.06 μSv). Also, the individual evaluation of the defined criteria was dose-independent. Conclusion: The adequate evaluation of hip dysplasia in children and young adults on pelvic radiographs is possible with reduced radiation dose, by simple using an exposure class of 800 instead of 400. … (more)
- Is Part Of:
- European journal of radiology. Volume 85:Issue 1(2016)
- Journal:
- European journal of radiology
- Issue:
- Volume 85:Issue 1(2016)
- Issue Display:
- Volume 85, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 85
- Issue:
- 1
- Issue Sort Value:
- 2016-0085-0001-0000
- Page Start:
- 233
- Page End:
- 238
- Publication Date:
- 2016-01
- Subjects:
- Pelvic X-ray examination -- Plain radiography -- Radiation dose reduction -- Quality management -- Radiation protection -- Developmental dysplasia of the hip
Medical radiology -- Periodicals
Radiology -- Periodicals
Radiologie médicale -- Périodiques
Medical radiology
Periodicals
616.075705 - Journal URLs:
- http://www.sciencedirect.com/science/journal/0720048X ↗
http://www.elsevier.com/homepage/elecserv.htt ↗
http://www.clinicalkey.com/dura/browse/journalIssue/0720048X ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/0720048X ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ejrad.2015.11.018 ↗
- Languages:
- English
- ISSNs:
- 0720-048X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.738050
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7508.xml