Characterization of the Exradin A26 microchamber for small field dosimetry. (September 2016)
- Record Type:
- Journal Article
- Title:
- Characterization of the Exradin A26 microchamber for small field dosimetry. (September 2016)
- Main Title:
- Characterization of the Exradin A26 microchamber for small field dosimetry
- Authors:
- Pasquino, M.
Cutaia, C.
Radici, L.
Gino, E.
Stasi, M. - Abstract:
- Abstract : Introduction: In modern radiation therapy accurate dosimetry within small photon fields is a challenge. The aim of the study is to investigate the main dosimetric characteristics of a new small field detector. Material and methods: An Exradin A26 microionization chamber was tested. Short term stability, dose linearity, polarity effect, angular dependence and energy dependence were evaluated for field sizes up to 10 × 10 cm 2 . Results were compared with those obtained with an Exradin A1SL IC. PDD curves, OAR profiles and OF values were measured for different field sizes. The results were compared with those obtained with an IC and an EDGE diode. Results: A26 IC readings showed a 0.2% and 0.1% relative standard deviation for 3 × 3 and 10 × 10 cm 2 fields, respectively. Dose response was linear to within 0.5%. The microchamber showed the maximum variation from the reference condition at the lowest dose rate. The polarity effect is within 0.5% for field size greater than 2 × 2 cm 2 ; for field size less than 2 × 2 cm 2 the effect increases, still remaining within 1%. No energy and angular dependence were observed. A26 IC PDD measurements agreed with those obtained with the EDGE diode. FWHM values show a better agreement with the EDGE diode values than A1SL. The penumbra values were greater than those obtained with the EDGE diode. Of values showed a good agreement with EDGE diode ones. Conclusions: The study confirms A26 IC as a reference detector for PDD and OARAbstract : Introduction: In modern radiation therapy accurate dosimetry within small photon fields is a challenge. The aim of the study is to investigate the main dosimetric characteristics of a new small field detector. Material and methods: An Exradin A26 microionization chamber was tested. Short term stability, dose linearity, polarity effect, angular dependence and energy dependence were evaluated for field sizes up to 10 × 10 cm 2 . Results were compared with those obtained with an Exradin A1SL IC. PDD curves, OAR profiles and OF values were measured for different field sizes. The results were compared with those obtained with an IC and an EDGE diode. Results: A26 IC readings showed a 0.2% and 0.1% relative standard deviation for 3 × 3 and 10 × 10 cm 2 fields, respectively. Dose response was linear to within 0.5%. The microchamber showed the maximum variation from the reference condition at the lowest dose rate. The polarity effect is within 0.5% for field size greater than 2 × 2 cm 2 ; for field size less than 2 × 2 cm 2 the effect increases, still remaining within 1%. No energy and angular dependence were observed. A26 IC PDD measurements agreed with those obtained with the EDGE diode. FWHM values show a better agreement with the EDGE diode values than A1SL. The penumbra values were greater than those obtained with the EDGE diode. Of values showed a good agreement with EDGE diode ones. Conclusions: The study confirms A26 IC as a reference detector for PDD and OAR measurements. … (more)
- Is Part Of:
- Physica medica. Volume 32(2016)Supplement 3
- Journal:
- Physica medica
- Issue:
- Volume 32(2016)Supplement 3
- Issue Display:
- Volume 32, Issue 3 (2016)
- Year:
- 2016
- Volume:
- 32
- Issue:
- 3
- Issue Sort Value:
- 2016-0032-0003-0000
- Page Start:
- 201
- Page End:
- Publication Date:
- 2016-09
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2016.07.682 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: