Dosimetric characterization of a commercial 2-D scintillation detector for quality assurance tests in scanned proton and carbon ion beams. (September 2016)
- Record Type:
- Journal Article
- Title:
- Dosimetric characterization of a commercial 2-D scintillation detector for quality assurance tests in scanned proton and carbon ion beams. (September 2016)
- Main Title:
- Dosimetric characterization of a commercial 2-D scintillation detector for quality assurance tests in scanned proton and carbon ion beams
- Authors:
- Russo, S.
Boi, D.
Mirandola, A.
Molinelli, S.
Mairani, A.
Mastella, E.
Magro, G.
Giordanengo, S.
Ciocca, M. - Abstract:
- Abstract : Introduction: Pencil beam scanning technique used at CNAO requires beam characteristics to be carefully and periodically checked to guarantee patient safety. Radiochromic films are used within daily QA checks. Purpose: This study aimed at characterizing the Lynx PT system (IBA Dosimetry) and assessing its suitability for QA tests, for proton and carbon ion beams. Materials and methods: The Lynx PT is a 2D high-resolution dosimetry system consisting of a scintillating screen coupled with a CCD camera, in a compact light-tight box. A dedicated software is used for image acquisition, raw data correction and fast data analysis. Lynx was characterized in terms of short-term stability, image geometrical distortion, uniformity, signal proportionality, response dependence on IRIS and on particle intensity, particle energy and type. Irradiations were performed at the Centro Nazionale di Adroterapia Oncologica (CNAO) for both protons and carbon ions. Measurements with Lynx were compared to those made with EBT3 reference films. The detector response dependence on radiation LET was also investigated. Results: Lynx basic characterization results have shown a proportional behavior with delivered number of particles. Strong dependence of the acquired signal on beam energy and particle type was found. No significant differences in the investigated beam parameters measured with Lynx and EBT3 film were observed. The detector also showed good performances in term of image uniformityAbstract : Introduction: Pencil beam scanning technique used at CNAO requires beam characteristics to be carefully and periodically checked to guarantee patient safety. Radiochromic films are used within daily QA checks. Purpose: This study aimed at characterizing the Lynx PT system (IBA Dosimetry) and assessing its suitability for QA tests, for proton and carbon ion beams. Materials and methods: The Lynx PT is a 2D high-resolution dosimetry system consisting of a scintillating screen coupled with a CCD camera, in a compact light-tight box. A dedicated software is used for image acquisition, raw data correction and fast data analysis. Lynx was characterized in terms of short-term stability, image geometrical distortion, uniformity, signal proportionality, response dependence on IRIS and on particle intensity, particle energy and type. Irradiations were performed at the Centro Nazionale di Adroterapia Oncologica (CNAO) for both protons and carbon ions. Measurements with Lynx were compared to those made with EBT3 reference films. The detector response dependence on radiation LET was also investigated. Results: Lynx basic characterization results have shown a proportional behavior with delivered number of particles. Strong dependence of the acquired signal on beam energy and particle type was found. No significant differences in the investigated beam parameters measured with Lynx and EBT3 film were observed. The detector also showed good performances in term of image uniformity and geometrical distortion. Conclusion: The Lynx PT system appears as a promising tool for measurements for both single pencil beam and scanned beam, enabling fast QA and beam optics optimization checks. … (more)
- Is Part Of:
- Physica medica. Volume 32(2016)Supplement 3
- Journal:
- Physica medica
- Issue:
- Volume 32(2016)Supplement 3
- Issue Display:
- Volume 32, Issue 3 (2016)
- Year:
- 2016
- Volume:
- 32
- Issue:
- 3
- Issue Sort Value:
- 2016-0032-0003-0000
- Page Start:
- 199
- Page End:
- Publication Date:
- 2016-09
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2016.07.675 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7456.xml