Evaluation of the relationship of corneal biomechanical metrics with physical intraocular pressure and central corneal thickness in ex vivo rabbit eye globes. (August 2015)
- Record Type:
- Journal Article
- Title:
- Evaluation of the relationship of corneal biomechanical metrics with physical intraocular pressure and central corneal thickness in ex vivo rabbit eye globes. (August 2015)
- Main Title:
- Evaluation of the relationship of corneal biomechanical metrics with physical intraocular pressure and central corneal thickness in ex vivo rabbit eye globes
- Authors:
- Bao, FangJun
Deng, ManLi
Wang, QinMei
Huang, JinHai
Yang, Jing
Whitford, Charles
Geraghty, Brendan
Yu, AYong
Elsheikh, Ahmed - Abstract:
- Abstract: The relationship of corneal biomechanical metrics provided by the Ocular Response Analyzer (ORA) and Corvis ST (CVS) with physical intraocular pressure (IOPp) and central corneal thickness (CCT) was evaluated. Thirty fresh enucleated eyes of 30 rabbits were used in ex vivo whole globe inflation experiments. IOPp was measured with a pressure transducer and increased from 7.5 to 37.5 mmHg in steps of 7.5 mmHg while biomechanical data was acquired using the ORA and CVS. At least 3 examinations were performed at each pressure level, where CCT and twelve biomechanical metrics were recorded and analyzed as a function of IOPp. The biomechanical metrics included corneal hysteresis (CH) and corneal resistance factor (CRF), obtained by the ORA. They also included the applanation times (A1T, A2T), lengths (A1L, A2L) and velocities (A1V, A2V), in addition to the highest concavity time (HCT), peak distance (PD), radius (HR) and deformation amplitude (DA), obtained by the CVS. The variation of CCT and the twelve biomechanical metrics for the 30 rabbit eyes tested across the 5 pressure stages considered (inter-pressure differences) were statistically significant (P = 0.00). IOPp was highly to moderately correlated with most biomechanical metrics, especially CRF, A1T, A1V, A2V, PD and DA, while the relationships with CH, A2T, A1L and HCT were poor. IOP has important influences on most corneal biomechanical metrics provided by CVS and ORA. Two biomechanical metrics A1V and HR wereAbstract: The relationship of corneal biomechanical metrics provided by the Ocular Response Analyzer (ORA) and Corvis ST (CVS) with physical intraocular pressure (IOPp) and central corneal thickness (CCT) was evaluated. Thirty fresh enucleated eyes of 30 rabbits were used in ex vivo whole globe inflation experiments. IOPp was measured with a pressure transducer and increased from 7.5 to 37.5 mmHg in steps of 7.5 mmHg while biomechanical data was acquired using the ORA and CVS. At least 3 examinations were performed at each pressure level, where CCT and twelve biomechanical metrics were recorded and analyzed as a function of IOPp. The biomechanical metrics included corneal hysteresis (CH) and corneal resistance factor (CRF), obtained by the ORA. They also included the applanation times (A1T, A2T), lengths (A1L, A2L) and velocities (A1V, A2V), in addition to the highest concavity time (HCT), peak distance (PD), radius (HR) and deformation amplitude (DA), obtained by the CVS. The variation of CCT and the twelve biomechanical metrics for the 30 rabbit eyes tested across the 5 pressure stages considered (inter-pressure differences) were statistically significant (P = 0.00). IOPp was highly to moderately correlated with most biomechanical metrics, especially CRF, A1T, A1V, A2V, PD and DA, while the relationships with CH, A2T, A1L and HCT were poor. IOP has important influences on most corneal biomechanical metrics provided by CVS and ORA. Two biomechanical metrics A1V and HR were influenced by CCT after correcting for the effect of IOP in most pressure stages, while the correlation with others were weak. Comparisons of research groups based on ORA and CVS with different IOPs and CCTs may lead to possible misinterpretations if both or one of which are not considered in the analysis. Highlights: 12 corneal biomechanical metrics were acquired in 5 different pressure stages. Intraocular pressure has significant influences on corneal biomechanical metrics. Central corneal thickness has less influences on corneal biomechanical metrics. … (more)
- Is Part Of:
- Experimental eye research. Volume 137(2015:Aug.)
- Journal:
- Experimental eye research
- Issue:
- Volume 137(2015:Aug.)
- Issue Display:
- Volume 137 (2015)
- Year:
- 2015
- Volume:
- 137
- Issue Sort Value:
- 2015-0137-0000-0000
- Page Start:
- 11
- Page End:
- 17
- Publication Date:
- 2015-08
- Subjects:
- Corneal biomechanical behavior -- Ex vivo -- Whole globe inflation experiment -- Intraocular pressure -- Central corneal thickness
Ophthalmology -- Periodicals
Eye -- Periodicals
Œil -- Périodiques
Ophthalmology
Periodicals
Electronic journals
612.8405 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00144835 ↗
http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=0014-4835;screen=info;ECOIP ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.exer.2015.05.018 ↗
- Languages:
- English
- ISSNs:
- 0014-4835
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3839.150000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7398.xml