Measurements and simulation of RF heating of implanted stereo‐electroencephalography electrodes during MR scans. Issue 4 (21st February 2018)
- Record Type:
- Journal Article
- Title:
- Measurements and simulation of RF heating of implanted stereo‐electroencephalography electrodes during MR scans. Issue 4 (21st February 2018)
- Main Title:
- Measurements and simulation of RF heating of implanted stereo‐electroencephalography electrodes during MR scans
- Authors:
- Bhusal, Bhumi
Bhattacharyya, Pallab
Baig, Tanvir
Jones, Stephen
Martens, Michael - Abstract:
- Abstract : Purpose: To assess RF‐induced heating during MRI of patients with implanted stereo‐electroencephalography electrodes. Methods: Simulations and experimental measurements using phantom and a head‐only transmit/receive coil on a 3T MR system were performed to evaluate temperature increases at the tip of an 8‐contact stereo‐electroencephalography electrode and an insulated wire partially immersed into the phantom. The lengths of wire producing maximum (resonant condition) and minimum (anti‐resonant condition) heating were evaluated for different entry modes and penetration depths. Results: For both wire and stereo‐electroencephalography electrode, resonant lengths were close to odd integral multiples of RF quarter wavelength in air and antiresonant length close to even integral multiples of RF quarter wavelength, both being unaffected by the entry mode. In the resonant condition, temperature increased by as much as a factor of 10 higher than that at antiresonant condition. Larger penetration depths did not change resonant length, but did lead to increased RF heating. Conclusion: For the partially immersed implants like stereo‐electroencephalography electrode, the resonant lengths were found to be independent of the penetration depths and entry modes, although the temperature increases may vary. Avoiding such lengths of cables can reduce the risk of tissue heating during in vivo MRI.
- Is Part Of:
- Magnetic resonance in medicine. Volume 80:Issue 4(2018)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 80:Issue 4(2018)
- Issue Display:
- Volume 80, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 80
- Issue:
- 4
- Issue Sort Value:
- 2018-0080-0004-0000
- Page Start:
- 1676
- Page End:
- 1685
- Publication Date:
- 2018-02-21
- Subjects:
- FDTD simulations -- functional MRI -- implant -- MR safety -- RF heating -- SEEG
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.27144 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7304.xml