Comparison of handheld ultrasound and automated breast ultrasound in women recalled after mammography screening. (May 2017)
- Record Type:
- Journal Article
- Title:
- Comparison of handheld ultrasound and automated breast ultrasound in women recalled after mammography screening. (May 2017)
- Main Title:
- Comparison of handheld ultrasound and automated breast ultrasound in women recalled after mammography screening
- Authors:
- Hellgren, Roxanna
Dickman, Paul
Leifland, Karin
Saracco, Ariel
Hall, Per
Celebioglu, Fuat - Abstract:
- Background: Automated breast volume scanner (ABVS) is an ultrasound (US) device with a wide scanner that sweeps over a large area of the breast and the acquired transverse images are sent to a workstation for reconstruction and review. Whether ABVS is as reliable as handheld US is, however, still not established. Purpose: To compare the sensitivity and specificity of ABVS to handheld breast US for detection of breast cancer, in the situation of recall after mammography screening. Material and Methods: A total of 113 women, five with bilateral suspicious findings, undergoing handheld breast US due to a suspicious mammographic finding in screening, underwent additional ABVS. The methods were assessed for each breast and each detected lesion separately and classified into two categories: breasts with mammographic suspicion of malignancy and breasts with a negative mammogram. Results: Twenty-six cancers were found in 25 women. In the category of breasts with a suspicious mammographic finding (n = 118), the sensitivity of both handheld US and ABVS was 88% (22/25). The specificity of handheld US was 93.5% (87/93) and ABVS was 89.2% (83/93). In the category of breasts with a negative mammography (n = 103), the sensitivity of handheld US and ABVS was 100% (1/1). The specificity of handheld US was 100% (102/102) and ABVS was 94.1% (96/102). Conclusion: ABVS can potentially replace handheld US in the investigation of women recalled from mammography screening due to a suspiciousBackground: Automated breast volume scanner (ABVS) is an ultrasound (US) device with a wide scanner that sweeps over a large area of the breast and the acquired transverse images are sent to a workstation for reconstruction and review. Whether ABVS is as reliable as handheld US is, however, still not established. Purpose: To compare the sensitivity and specificity of ABVS to handheld breast US for detection of breast cancer, in the situation of recall after mammography screening. Material and Methods: A total of 113 women, five with bilateral suspicious findings, undergoing handheld breast US due to a suspicious mammographic finding in screening, underwent additional ABVS. The methods were assessed for each breast and each detected lesion separately and classified into two categories: breasts with mammographic suspicion of malignancy and breasts with a negative mammogram. Results: Twenty-six cancers were found in 25 women. In the category of breasts with a suspicious mammographic finding (n = 118), the sensitivity of both handheld US and ABVS was 88% (22/25). The specificity of handheld US was 93.5% (87/93) and ABVS was 89.2% (83/93). In the category of breasts with a negative mammography (n = 103), the sensitivity of handheld US and ABVS was 100% (1/1). The specificity of handheld US was 100% (102/102) and ABVS was 94.1% (96/102). Conclusion: ABVS can potentially replace handheld US in the investigation of women recalled from mammography screening due to a suspicious finding. Due to the small size of our study population, further investigation with larger study populations is necessary before the implementation of such practice. … (more)
- Is Part Of:
- Acta radiologica. Volume 58:Number 5(2017:May)
- Journal:
- Acta radiologica
- Issue:
- Volume 58:Number 5(2017:May)
- Issue Display:
- Volume 58, Issue 5 (2017)
- Year:
- 2017
- Volume:
- 58
- Issue:
- 5
- Issue Sort Value:
- 2017-0058-0005-0000
- Page Start:
- 515
- Page End:
- 520
- Publication Date:
- 2017-05
- Subjects:
- Breast -- ultrasound -- primary neoplasms -- mammography
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1177/0284185116665421 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7285.xml