Development of a Double-Lumen Cannula for a Percutaneous RVAD. Issue 4 (July 2015)
- Record Type:
- Journal Article
- Title:
- Development of a Double-Lumen Cannula for a Percutaneous RVAD. Issue 4 (July 2015)
- Main Title:
- Development of a Double-Lumen Cannula for a Percutaneous RVAD
- Authors:
- Wang, Dongfang
Jones, Cameron
Ballard-Croft, Cherry
Zhao, Ju
Zhao, Guangfeng
Topaz, Stephen
Zwischenberger, Joseph B. - Abstract:
- Abstract : The objectives were to design/fabricate a double-lumen cannula (DLC) for a percutaneous right ventricular assist device (pRVAD) and to test the feasibility/performance of this pRVAD system. A 27 Fr DLC prototype was made and tested in six adult sheep. The pRVAD DLC was inserted into the right jugular vein; advanced through the superior vena cava, the right atrium (RA), the right ventricle (RV); ending in the pulmonary artery (PA). A CentriMag pump and optional gas exchanger were connected to the DLC. Blood was withdrawn from RA, pumped through gas exchanger, and perfused PA. Maximal pumping flow was maintained for 2 hours. The pRVAD DLC was successfully deployed in all six sheep. In first three sheep, maximal average pumping flow was less than 3 L/min because the DLC was advanced too far with drainage opening against RA side wall. In last three sheep with well-positioned DLC, average maximal flow was more than 3.5 L/min. The gas exchanger provided up to 230 ml/min CO2 removal and 174 ml/min O2 transfer. Our DLC-based pRVAD system is feasible for percutaneous right heart and respiratory assistance through a single cannulation. The pRVAD DLC can be easily placed prophylactically during left ventricular assist device implantation and removed as needed without additional open chest procedures.
- Is Part Of:
- ASAIO journal. Volume 61:Issue 4(2015)
- Journal:
- ASAIO journal
- Issue:
- Volume 61:Issue 4(2015)
- Issue Display:
- Volume 61, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 61
- Issue:
- 4
- Issue Sort Value:
- 2015-0061-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-07
- Subjects:
- right ventricular assist device -- double-lumen cannula -- percutaneous -- sheep
Artificial organs -- Periodicals
617 - Journal URLs:
- http://journals.lww.com/asaiojournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAT.0000000000000227 ↗
- Languages:
- English
- ISSNs:
- 1058-2916
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1738.840500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 7181.xml