A compliant aortic model for in vitro simulations: Design and manufacturing process. (September 2018)
- Record Type:
- Journal Article
- Title:
- A compliant aortic model for in vitro simulations: Design and manufacturing process. (September 2018)
- Main Title:
- A compliant aortic model for in vitro simulations: Design and manufacturing process
- Authors:
- Marconi, Stefania
Lanzarone, Ettore
van Bogerijen, Guido H.W.
Conti, Michele
Secchi, Francesco
Trimarchi, Santi
Auricchio, Ferdinando - Abstract:
- Highlights: A compliant silicone model of the entire human aorta is proposed. The model mimics both geometrical and mechanical local properties of the aorta. It can be used as a customizable tool for the in vitro testing of vascular devices. Graphical abstract: Abstract: We design and manufacture a silicone model of the human aorta, able to mimic both the geometrical and the mechanical properties of physiological individuals, with a specific focus on reproducing the compliance. In fact, while the models available in the literature exhibit an unrealistic compliant behavior, though they are detailed from the geometrical viewpoint, here the goal is to provide an accurate compliant tool for in vitro testing the devices that interface with the vascular system. A parametric design of the aortic model is obtained based on the available literature data, and the model is manufactured with a specific silicone mixture using rapid prototyping and molding techniques. The manufactured prototype has been tested by means of computed tomography scans for evaluating the matching of the mechanical properties with the desired ones. Results show a high degree of adherence between the imposed and the measured compliance values for each main aortic section. Thus, our work proves the feasibility of the approach, and the possibility to manufacture compliant models that reproduce the mechanical behavior of the aorta for in vitro studies.
- Is Part Of:
- Medical engineering & physics. Volume 59(2018)
- Journal:
- Medical engineering & physics
- Issue:
- Volume 59(2018)
- Issue Display:
- Volume 59, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 59
- Issue:
- 2018
- Issue Sort Value:
- 2018-0059-2018-0000
- Page Start:
- 21
- Page End:
- 29
- Publication Date:
- 2018-09
- Subjects:
- Aortic model -- Compliance reproduction -- Rapid prototyping -- Parametric design -- Molding -- In vitro simulation
Biomedical engineering -- Periodicals
Biomedical Engineering -- Periodicals
Physics -- Periodicals
Génie biomédical -- Périodiques
Biomedical engineering
Electronic journals
Periodicals
610.28 - Journal URLs:
- http://www.medengphys.com ↗
http://www.sciencedirect.com/science/journal/13504533 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/13504533 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/13504533 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.medengphy.2018.04.022 ↗
- Languages:
- English
- ISSNs:
- 1350-4533
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5527.323000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7183.xml