Validation of frame-based positioning accuracy with cone-beam computed tomography in Gamma Knife Icon radiosurgery. (August 2018)
- Record Type:
- Journal Article
- Title:
- Validation of frame-based positioning accuracy with cone-beam computed tomography in Gamma Knife Icon radiosurgery. (August 2018)
- Main Title:
- Validation of frame-based positioning accuracy with cone-beam computed tomography in Gamma Knife Icon radiosurgery
- Authors:
- Stieler, Florian
Wenz, Frederik
Schweizer, Beate
Polednik, Martin
Giordano, Frank Anton
Mai, Sabine - Abstract:
- Highlights: Frame accuracy with 4 posts, validated with CBCT, is expectedly high. Phantom measurements showed similar deviations if the left or right post is removed. For clinical situations with 3 posts, position verification is suggested. If CBCT is not availible, the P-CT/MR should be repeated. Abstract: Background: Frame based positioning accuracy in Gamma Knife (GK) stereotactic radiosurgery (SRS) is extremely high but removal of a post may be necessary to enable the treatment in selected patients. Objective: To verify the positioning accuracy in clinical scenarios with 4 and 3 posts in patients and phantoms using cone-beam CT (CBCT) of Gamma Knife Icon™. Methods: We analyzed positioning accuracy for 12 patients with standard 4 post setup using pretreatment CBCT (pre-CBCT) on GK Icon™ and report 4 patients with different clinical scenarios (removal of a post). We performed phantom measurements to verify the frame accuracy via CBCT in different clinical scenarios without the influence of the human patient. Results: Mean frame accuracy for 12 patients with 4 posts was 0.35 mm/0.34 degree. Mean motion during treatment was 0.11 mm/0.04 degree. For two of the clinical scenarios where a post was removed, we found acceptable deviations within 0.66 mm/0.61 degree. For 2 patients, a deviation of 2.94 mm/−3.47 degree and 1.85 mm/−0.74 degree was found and replanning was necessary. Phantom measurements showed good agreement when planning MR/CT was performed with 4 or 3 post.Highlights: Frame accuracy with 4 posts, validated with CBCT, is expectedly high. Phantom measurements showed similar deviations if the left or right post is removed. For clinical situations with 3 posts, position verification is suggested. If CBCT is not availible, the P-CT/MR should be repeated. Abstract: Background: Frame based positioning accuracy in Gamma Knife (GK) stereotactic radiosurgery (SRS) is extremely high but removal of a post may be necessary to enable the treatment in selected patients. Objective: To verify the positioning accuracy in clinical scenarios with 4 and 3 posts in patients and phantoms using cone-beam CT (CBCT) of Gamma Knife Icon™. Methods: We analyzed positioning accuracy for 12 patients with standard 4 post setup using pretreatment CBCT (pre-CBCT) on GK Icon™ and report 4 patients with different clinical scenarios (removal of a post). We performed phantom measurements to verify the frame accuracy via CBCT in different clinical scenarios without the influence of the human patient. Results: Mean frame accuracy for 12 patients with 4 posts was 0.35 mm/0.34 degree. Mean motion during treatment was 0.11 mm/0.04 degree. For two of the clinical scenarios where a post was removed, we found acceptable deviations within 0.66 mm/0.61 degree. For 2 patients, a deviation of 2.94 mm/−3.47 degree and 1.85 mm/−0.74 degree was found and replanning was necessary. Phantom measurements showed good agreement when planning MR/CT was performed with 4 or 3 post. Larger deviations of 0.86 mm/0.88 degree were detected when a post was removed after planning MR/CT. Conclusion: The frame accuracy with 4 posts before and during GK treatments is as high as expected. For clinical situations, where a post is removed after planning-CT/MR, pre-treatment position verification is strongly suggested using stereotactic CBCT or the P-CT/MR should be repeated to avoid possible mistreatments. … (more)
- Is Part Of:
- Physica medica. Volume 52(2018)
- Journal:
- Physica medica
- Issue:
- Volume 52(2018)
- Issue Display:
- Volume 52, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 52
- Issue:
- 2018
- Issue Sort Value:
- 2018-0052-2018-0000
- Page Start:
- 93
- Page End:
- 97
- Publication Date:
- 2018-08
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.06.640 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7162.xml