The impacts of a new electrochemical etch cycle for the Public Health England neutron personal dosimetry service. (November 2017)
- Record Type:
- Journal Article
- Title:
- The impacts of a new electrochemical etch cycle for the Public Health England neutron personal dosimetry service. (November 2017)
- Main Title:
- The impacts of a new electrochemical etch cycle for the Public Health England neutron personal dosimetry service
- Authors:
- Hager, Luke
Tanner, Rick
Gilvin, Phil
Eakins, Jonathan
Baker, Sean - Abstract:
- Abstract: The Public Health England (PHE) neutron and radon personal dosimetry services based on electrochemical etching of PADC have been running since 1986. Changes to the detector material have required a new etch cycle to be developed to improve linearity at high doses: A new supplier of PADC was able to meet the stringent criteria for thickness, response and background, but it was found that the tracks were, on average, larger, resulting in increased track overlap, saturation of the signal at lower dose levels and a reduction in the maximum dose measurable. This dose linearity problem meant that a more severe correction was needed, which led to increased measurement uncertainty. A new etch cycle has reduced the mean track size, and reduced track overlap, meaning that the maximum reportable dose is increased and dose assessments at higher doses are more reliable. Also reported on here are the results of tests on the energy and angle dependences of response, and the performance of the dosemeter in workplace fields, with data compared from dosemeters etched with both the old and the new processes. The results show that the new etch cycle produces an improvement to the routine service. Highlights: Improved dose linearity with a new etch cycle. Energy dependence of response changes with a new etch cycle. Effect on workplace field normalization factor with a new etch cycle.
- Is Part Of:
- Radiation measurements. Volume 106(2017:Nov.)
- Journal:
- Radiation measurements
- Issue:
- Volume 106(2017:Nov.)
- Issue Display:
- Volume 106 (2017)
- Year:
- 2017
- Volume:
- 106
- Issue Sort Value:
- 2017-0106-0000-0000
- Page Start:
- 303
- Page End:
- 311
- Publication Date:
- 2017-11
- Subjects:
- PADC -- Neutrons -- Personal dosimetry -- Energy dependence of response -- Dose linearity -- Workplace field normalization factor
Nuclear emulsions -- Periodicals
Particle tracks (Nuclear physics) -- Periodicals
Thermoluminescence -- Periodicals
Cosmic rays -- Periodicals
Radiation -- Measurement -- Periodicals
Radiometry -- Periodicals
Radiation Monitoring -- Periodicals
Émulsions nucléaires -- Périodiques
Particules (Physique nucléaire) -- Traces -- Périodiques
Thermoluminescence -- Périodiques
Rayonnement cosmique -- Périodiques
Radiométrie -- Périodiques
539.77 - Journal URLs:
- http://www.sciencedirect.com/science/journal/13504487 ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiation-measurements/ ↗ - DOI:
- 10.1016/j.radmeas.2017.05.010 ↗
- Languages:
- English
- ISSNs:
- 1350-4487
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7227.973000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7160.xml