An international collaborative study to compare different von Willebrand factor glycoprotein Ib binding activity assays: the COMPASS‐VWF study. (19th July 2018)
- Record Type:
- Journal Article
- Title:
- An international collaborative study to compare different von Willebrand factor glycoprotein Ib binding activity assays: the COMPASS‐VWF study. (19th July 2018)
- Main Title:
- An international collaborative study to compare different von Willebrand factor glycoprotein Ib binding activity assays: the COMPASS‐VWF study
- Authors:
- Szederjesi, A.
Baronciani, L.
Budde, U.
Castaman, G.
Lawrie, A. S.
Liu, Y.
Montgomery, R.
Peyvandi, F.
Schneppenheim, R.
Várkonyi, A.
Patzke, J.
Bodó, I. - Abstract:
- Abstract : Essentials New VWF activity assays are increasingly used but information on their comparability is limited. This is an ISTH SSC‐organized study (expert labs, 5 countries) to compare all available assays. VWF activity by six assays correlated well with each other. The new assays show improved characteristics ‐ minor differences are noted. Summary: Background: Several new assays have become available to measure von Willebrand factor (VWF) activity. The new assays appear to have improved performance characteristics compared with the old reference standard, ristocetin cofactor activity (VWF:RCo), but information is limited about how they compare with VWF:RCo and each other. Methods: The von Willebrand factor Subcommittee of the International Society for Thrombosis and Haemostasis (ISTH) Scientific and Standardization Committee (SSC) designed a collaborative study involving expert laboratories from several countries to compare available tests with each other and with VWF:RCo. Eight laboratories from five countries were provided with blinded samples from normal healthy individuals and well‐characterized clinical cases. Laboratories measured VWF activity using all tests available to them; data from six laboratories, not affected by thawing during transportation, are included in this study. Results: All tests correlated well with VWF:RCo activity ( r ‐values ranged from 0.963 to 0.989). Slightly steeper regression lines for VWF:Ab and VWF:GPIbM were clinicallyAbstract : Essentials New VWF activity assays are increasingly used but information on their comparability is limited. This is an ISTH SSC‐organized study (expert labs, 5 countries) to compare all available assays. VWF activity by six assays correlated well with each other. The new assays show improved characteristics ‐ minor differences are noted. Summary: Background: Several new assays have become available to measure von Willebrand factor (VWF) activity. The new assays appear to have improved performance characteristics compared with the old reference standard, ristocetin cofactor activity (VWF:RCo), but information is limited about how they compare with VWF:RCo and each other. Methods: The von Willebrand factor Subcommittee of the International Society for Thrombosis and Haemostasis (ISTH) Scientific and Standardization Committee (SSC) designed a collaborative study involving expert laboratories from several countries to compare available tests with each other and with VWF:RCo. Eight laboratories from five countries were provided with blinded samples from normal healthy individuals and well‐characterized clinical cases. Laboratories measured VWF activity using all tests available to them; data from six laboratories, not affected by thawing during transportation, are included in this study. Results: All tests correlated well with VWF:RCo activity ( r ‐values ranged from 0.963 to 0.989). Slightly steeper regression lines for VWF:Ab and VWF:GPIbM were clinically insignificant. The new assays showed improved performance characteristics. Of the commercially available assays, the VWF:GPIbR using the AcuStar system was the most sensitive and could reliably detect VWF activity below 1 IU dL −1 . The lower limit of the measuring interval for the VWF:GPIbM and the VWF:GPIbR assays was in the 3–4 and 3–6 IU dL −1 range, respectively. Inter‐laboratory variation was also improved for most new assays. Conclusion: All VWF activity assays correlated well with each other and the VWF:RCo assay. The slight differences in characteristics found in the COMPASS‐VWF study will assist the VWF community in interpreting and comparing activity results. … (more)
- Is Part Of:
- Journal of thrombosis and haemostasis. Volume 16:Number 8(2018)
- Journal:
- Journal of thrombosis and haemostasis
- Issue:
- Volume 16:Number 8(2018)
- Issue Display:
- Volume 16, Issue 8 (2018)
- Year:
- 2018
- Volume:
- 16
- Issue:
- 8
- Issue Sort Value:
- 2018-0016-0008-0000
- Page Start:
- 1604
- Page End:
- 1613
- Publication Date:
- 2018-07-19
- Subjects:
- clinical laboratory techniques -- hemorrhagic disorders -- hemostasis -- von Willebrand disease -- von Willebrand factor
Thrombosis -- Periodicals
Hemostasis -- Periodicals
Blood coagulation disorders -- Periodicals
616.1 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1538-7836 ↗
http://www.blackwellpublishing.com/journals/jth ↗
https://www.sciencedirect.com/journal/journal-of-thrombosis-and-haemostasis ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/jth.14206 ↗
- Languages:
- English
- ISSNs:
- 1538-7933
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5069.345000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 7121.xml