Evaluation of the Virclia® automated chemiluminescent immunoassay system for diagnosing pneumonia caused by Mycoplasma pneumoniae. Issue 6 (6th March 2018)
- Record Type:
- Journal Article
- Title:
- Evaluation of the Virclia® automated chemiluminescent immunoassay system for diagnosing pneumonia caused by Mycoplasma pneumoniae. Issue 6 (6th March 2018)
- Main Title:
- Evaluation of the Virclia® automated chemiluminescent immunoassay system for diagnosing pneumonia caused by Mycoplasma pneumoniae
- Authors:
- Ortiz de la Tabla, Victoria
Berruezo, Marta
García Payá, Elena
Fernández, Marta
García, José Alberto
Masiá, Mar
Gutiérrez, Félix - Abstract:
- Abstract : Background: Mycoplasma pneumoniae is considered an important etiologic agent of community‐acquired pneumonia (CAP) in outpatients. We aimed to evaluate the diagnostic accuracy of a quick automated chemiluminescent immunoassay (CLIA) for M. pneumoniae in a population‐based prospective study of CAP. Methods: A total of 137 outpatients diagnosed with CAP were included in the study. Acute‐ and convalescent phase sera were analyzed for IgG and IgM to M. pneumoniae with both CLIA (VirClia ® ) and ELISA immunoassays. Conventional serological criteria by quantitative ELISA were considered as reference standard. Sensitivity and specificity of the assay were assessed with the construction of receiver operating characteristic (ROC) curves, and the kappa index was used to evaluate the accuracy of the IgG and IgM determinations in the acute phase. Results: Thirty‐eight patients were diagnosed with pneumonia by M. pneumoniae . ROC curves for IgG and IgM of convalescent and acute phase (C/A) quotients by the CLIA and ELISA assays were comparable. Specifically, for the CLIA, the best C/A quotient for IgG was 2.617 (sensitivity, 94.9%; specificity, 99.9%), and for IgM 1.400 (sensitivity, 65.8%; specificity, 100%). Regarding the acute phase, the best diagnostic accuracy for the CLIA was obtained with an IgG index of 1.120 (sensitivity, 89.5%; specificity, 73.7%). The CLIA was very simple to execute and required a minimum sample handling. Conclusion: The accuracy of the Virclia ®Abstract : Background: Mycoplasma pneumoniae is considered an important etiologic agent of community‐acquired pneumonia (CAP) in outpatients. We aimed to evaluate the diagnostic accuracy of a quick automated chemiluminescent immunoassay (CLIA) for M. pneumoniae in a population‐based prospective study of CAP. Methods: A total of 137 outpatients diagnosed with CAP were included in the study. Acute‐ and convalescent phase sera were analyzed for IgG and IgM to M. pneumoniae with both CLIA (VirClia ® ) and ELISA immunoassays. Conventional serological criteria by quantitative ELISA were considered as reference standard. Sensitivity and specificity of the assay were assessed with the construction of receiver operating characteristic (ROC) curves, and the kappa index was used to evaluate the accuracy of the IgG and IgM determinations in the acute phase. Results: Thirty‐eight patients were diagnosed with pneumonia by M. pneumoniae . ROC curves for IgG and IgM of convalescent and acute phase (C/A) quotients by the CLIA and ELISA assays were comparable. Specifically, for the CLIA, the best C/A quotient for IgG was 2.617 (sensitivity, 94.9%; specificity, 99.9%), and for IgM 1.400 (sensitivity, 65.8%; specificity, 100%). Regarding the acute phase, the best diagnostic accuracy for the CLIA was obtained with an IgG index of 1.120 (sensitivity, 89.5%; specificity, 73.7%). The CLIA was very simple to execute and required a minimum sample handling. Conclusion: The accuracy of the Virclia ® assay for the diagnosis of M. pneumoniae infection in outpatients with CAP was equivalent to the quantitative ELISA. The CLIA was quicker to perform and displayed better analytic workability than conventional ELISA. … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 32:Issue 6(2018)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 32:Issue 6(2018)
- Issue Display:
- Volume 32, Issue 6 (2018)
- Year:
- 2018
- Volume:
- 32
- Issue:
- 6
- Issue Sort Value:
- 2018-0032-0006-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2018-03-06
- Subjects:
- antibodies -- chemiluminescent immunoassay -- microbial diagnosis -- Mycoplasma pneumoniae -- pneumonia -- serology
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.22431 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7062.xml