Concordance between point‐of‐care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients. Issue 6 (3rd March 2018)
- Record Type:
- Journal Article
- Title:
- Concordance between point‐of‐care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients. Issue 6 (3rd March 2018)
- Main Title:
- Concordance between point‐of‐care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients
- Authors:
- Prakash, Shivesh
Bihari, Shailesh
Lim, Zhan Y.
Verghese, Santosh
Kulkarni, Hemant
Bersten, Andrew D. - Abstract:
- Abstract : Background: We tested the hypothesis that the results of the same test performed on point‐of‐care blood gas analysis (BGA) machine and automatic analyzer (AA) machine in central laboratory have high degree of concordance in critical care patients and that the two test methods could be used interchangeably. Methods: We analyzed 9398 matched pairs of BGA and AA results, obtained from 1765 patients. Concentration pairs of the following analytes were assessed: hemoglobin, glucose, sodium, potassium, chloride, and bicarbonate. We determined the agreement using concordance correlation coefficient (CCC) and Bland‐Altman analysis. The difference in results was also assessed against the United States Clinical Laboratory Improvement Amendments (US‐CLIA) 88 rules. The test results were considered to be interchangeable if they were within the US‐CLIA variability criteria and would not alter the clinical management when compared to each other. Results: The median time interval between sampling for BGA and AA in each result pair was 5 minutes. The CCC values ranged from 0.89(95% CI 0.89‐0.90) for chloride to 0.98(95% CI 0.98‐0.99) for hemoglobin. The largest bias was for hemoglobin. The limits of agreement relative to bias were largest for sodium, with 3.4% of readings outside the US‐CLIA variation rule. The number of readings outside the US‐CLIA acceptable variation was highest for glucose (7.1%) followed by hemoglobin (5.9%) and chloride (5.2%). Conclusion: We conclude thatAbstract : Background: We tested the hypothesis that the results of the same test performed on point‐of‐care blood gas analysis (BGA) machine and automatic analyzer (AA) machine in central laboratory have high degree of concordance in critical care patients and that the two test methods could be used interchangeably. Methods: We analyzed 9398 matched pairs of BGA and AA results, obtained from 1765 patients. Concentration pairs of the following analytes were assessed: hemoglobin, glucose, sodium, potassium, chloride, and bicarbonate. We determined the agreement using concordance correlation coefficient (CCC) and Bland‐Altman analysis. The difference in results was also assessed against the United States Clinical Laboratory Improvement Amendments (US‐CLIA) 88 rules. The test results were considered to be interchangeable if they were within the US‐CLIA variability criteria and would not alter the clinical management when compared to each other. Results: The median time interval between sampling for BGA and AA in each result pair was 5 minutes. The CCC values ranged from 0.89(95% CI 0.89‐0.90) for chloride to 0.98(95% CI 0.98‐0.99) for hemoglobin. The largest bias was for hemoglobin. The limits of agreement relative to bias were largest for sodium, with 3.4% of readings outside the US‐CLIA variation rule. The number of readings outside the US‐CLIA acceptable variation was highest for glucose (7.1%) followed by hemoglobin (5.9%) and chloride (5.2%). Conclusion: We conclude that there is moderate to substantial concordance between AA and BGA machines on tests performed in critically ill patients. However, the two tests methods cannot be used interchangeably, except for potassium. … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 32:Issue 6(2018)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 32:Issue 6(2018)
- Issue Display:
- Volume 32, Issue 6 (2018)
- Year:
- 2018
- Volume:
- 32
- Issue:
- 6
- Issue Sort Value:
- 2018-0032-0006-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2018-03-03
- Subjects:
- automatic analyzer -- blood gas analysis -- Electrolytes -- hemoglobin
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.22425 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7062.xml