Comparison of the effect of gel used in two different serum separator tubes for thyroid function tests. Issue 6 (7th March 2018)
- Record Type:
- Journal Article
- Title:
- Comparison of the effect of gel used in two different serum separator tubes for thyroid function tests. Issue 6 (7th March 2018)
- Main Title:
- Comparison of the effect of gel used in two different serum separator tubes for thyroid function tests
- Authors:
- Ercan, Müjgan
Fırat Oğuz, Esra
Akbulut, Emiş Deniz
Yilmaz, Mustafa
Turhan, Turan - Abstract:
- Abstract : Background: Selection and verification of blood collection tubes is an important preanalytical issue in clinical laboratories. Today, gel tubes are commonly used with many advantages, although they are known to cause interference in immunoassay methods. In this study, we aimed to compare SSTs of two different suppliers (Ayset clot activator & Gel and Becton Dickinson (BD) Vacutainer SST II advance) with reference tubes and evaluate the effect of storage time in terms of commonly used endocrine tests such as thyroid‐stimulating hormone (TSH), free thyroxine (fT4), and free triiodothyronine (fT3). Methods: Fifty‐five volunteers were included in the study. Samples were taken into three different tubes and analyzed for serum TSH, fT4, and fT3 on Architectci8200 Immunoassay System. Clinical decision levels were estimated using total allowable error (TEa). Results: No difference was found between tubes in terms of TSH, fT3, and fT4 levels. From a statistical standpoint, TSH and fT4 levels were no longer stable during 24, 48, and 72 hours storage time periods. However, their variations were not clinically significant. Conclusion: Ayset clot activator & Gel tubes and BD Vacutainer SST II advance tubes have comparable results with glass tube in terms of TSH, fT3, and fT4 levels on Architect ci8200 Immunoassay Systems. From a clinical standpoint, serum TSH, fT4, and fT3 concentrations may be considered as stable when storing these tubes over 72 hours.
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 32:Issue 6(2018)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 32:Issue 6(2018)
- Issue Display:
- Volume 32, Issue 6 (2018)
- Year:
- 2018
- Volume:
- 32
- Issue:
- 6
- Issue Sort Value:
- 2018-0032-0006-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2018-03-07
- Subjects:
- blood specimen collection -- gel separator tubes -- immunoassay -- preanalytical error -- thyroid function tests
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.22427 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7048.xml