Low-dose CT pulmonary angiography on a 15-year-old CT scanner: a feasibility study. (December 2016)
- Record Type:
- Journal Article
- Title:
- Low-dose CT pulmonary angiography on a 15-year-old CT scanner: a feasibility study. (December 2016)
- Main Title:
- Low-dose CT pulmonary angiography on a 15-year-old CT scanner: a feasibility study
- Authors:
- Kaup, Moritz
Gruber-Rouh, Tatjana
Scholtz, Jan E
Albrecht, Moritz H
Bucher, Andreas
Frellesen, Claudia
Vogl, Thomas J
Beeres, Martin - Abstract:
- Background: Computed tomography (CT) low-dose (LD) imaging is used to lower radiation exposure, especially in vascular imaging; in current literature, this is mostly on latest generation high-end CT systems. Purpose: To evaluate the effects of reduced tube current on objective and subjective image quality of a 15-year-old 16-slice CT system for pulmonary angiography (CTPA). Material and Methods: CTPA scans from 60 prospectively randomized patients (28 men, 32 women) were examined in this study on a 15-year-old 16-slice CT scanner system. Standard CT (SD) settings were 100 kV and 150 mAs, LD settings were 100 kV and 50 mAs. Attenuation of the pulmonary trunk, various anatomic landmarks, and image noise were quantitatively measured; contrast-to-noise ratios (CNR) and signal-to-noise ratios (SNR) were calculated. Three independent blinded radiologists subjectively rated each image series using a 5-point grading scale. Results: CT dose index (CTDI) in the LD series was 66.46% lower compared to the SD settings (2.49 ± 0.55 mGy versus 7.42 ± 1.17 mGy). Attenuation of the pulmonary trunk showed similar results for both series (SD 409.55 ± 91.04 HU; LD 380.43 HU ± 93.11 HU; P = 0.768). Subjective image analysis showed no significant differences between SD and LD settings regarding the suitability for detection of central and peripheral PE (central SD/LD, 4.88; intra-class correlation coefficients [ICC], 0.894/4.83; ICC, 0.745; peripheral SD/LD, 4.70; ICC, 0.943/4.57; ICC, 0.919;Background: Computed tomography (CT) low-dose (LD) imaging is used to lower radiation exposure, especially in vascular imaging; in current literature, this is mostly on latest generation high-end CT systems. Purpose: To evaluate the effects of reduced tube current on objective and subjective image quality of a 15-year-old 16-slice CT system for pulmonary angiography (CTPA). Material and Methods: CTPA scans from 60 prospectively randomized patients (28 men, 32 women) were examined in this study on a 15-year-old 16-slice CT scanner system. Standard CT (SD) settings were 100 kV and 150 mAs, LD settings were 100 kV and 50 mAs. Attenuation of the pulmonary trunk, various anatomic landmarks, and image noise were quantitatively measured; contrast-to-noise ratios (CNR) and signal-to-noise ratios (SNR) were calculated. Three independent blinded radiologists subjectively rated each image series using a 5-point grading scale. Results: CT dose index (CTDI) in the LD series was 66.46% lower compared to the SD settings (2.49 ± 0.55 mGy versus 7.42 ± 1.17 mGy). Attenuation of the pulmonary trunk showed similar results for both series (SD 409.55 ± 91.04 HU; LD 380.43 HU ± 93.11 HU; P = 0.768). Subjective image analysis showed no significant differences between SD and LD settings regarding the suitability for detection of central and peripheral PE (central SD/LD, 4.88; intra-class correlation coefficients [ICC], 0.894/4.83; ICC, 0.745; peripheral SD/LD, 4.70; ICC, 0.943/4.57; ICC, 0.919; all P > 0.4). Conclusion: The LD protocol, on a 15-year-old CT scanner system without current high-end hardware or post-processing tools, led to a dose reduction of approximately 67% with similar subjective image quality and delineation of central and peripheral pulmonary arteries. … (more)
- Is Part Of:
- Acta radiologica open. Volume 5:Number 12(2016)
- Journal:
- Acta radiologica open
- Issue:
- Volume 5:Number 12(2016)
- Issue Display:
- Volume 5, Issue 12 (2016)
- Year:
- 2016
- Volume:
- 5
- Issue:
- 12
- Issue Sort Value:
- 2016-0005-0012-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-12
- Subjects:
- Computed tomography pulmonary angiography (CTPA); low-dose imaging; CT pulmonary angiography; CT radiation exposure
Radiology -- Periodicals
Diagnostic Imaging -- Periodicals
Radiology
Periodicals
616.075705 - Journal URLs:
- http://arr.sagepub.com/ ↗
http://arr.sagepub.com/ ↗
http://www.uk.sagepub.com ↗
http://www.ncbi.nlm.nih.gov/pmc/journals/2704/ ↗ - DOI:
- 10.1177/2058460116684371 ↗
- Languages:
- English
- ISSNs:
- 2058-4601
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6980.xml