Comparison of corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and digital caliper in frozen canine corneas. (7th November 2017)
- Record Type:
- Journal Article
- Title:
- Comparison of corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and digital caliper in frozen canine corneas. (7th November 2017)
- Main Title:
- Comparison of corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and digital caliper in frozen canine corneas
- Authors:
- Jeong, Seowoo
Kang, Seonmi
Park, Sangwan
Park, Eunjin
Lim, Jaegook
Nam, Taekjin
Seo, Kangmoon - Abstract:
- Abstract: Objective: To evaluate the clinical efficacy among ultrasound pachymetry (UP), ultrasound biomicroscopy (UBM), and manual measurement (MM) with digital caliper by comparing corneal thickness (COT) values obtained from frozen canine corneas ex vivo . Procedures: COT was measured using UP, UBM, and MM in eight enucleated normal canine eyes frozen at −20 °C for 4 weeks. After thawing at room temperature for 2 h, the COT values were obtained from five sites on each cornea: central corneal thickness (CCOT), thickness at the 3, 6, 9, and 12 o'clock positions (3COT, 6COT, 9COT, and 12COT, respectively). For each device, measurements were performed three times at each designated site by one operator. Intraclass correlation coefficient, repeated‐measures analysis of variance, and Bland–Altman plot were performed for statistical analyses. Results: The mean CCOT was 839.0 ± 138.2, 857.6 ± 127.9, and 849.1 ± 132.8 μm for UP, UBM, and MM, respectively. There was no statistically significant difference among the measurements by all three devices ( P > 0.05). The statistical agreement among these devices was also shown to be good according to Bland–Altman plots. In addition, the values measured on the peripheral sites of the cornea (3COT, 6COT, 9COT, and 12COT) also revealed no significant differences among the three devices. The difference between CCOT and peripheral corneal thickness was also statistically insignificant. Conclusions: The UP, UBM, and MM measurements showedAbstract: Objective: To evaluate the clinical efficacy among ultrasound pachymetry (UP), ultrasound biomicroscopy (UBM), and manual measurement (MM) with digital caliper by comparing corneal thickness (COT) values obtained from frozen canine corneas ex vivo . Procedures: COT was measured using UP, UBM, and MM in eight enucleated normal canine eyes frozen at −20 °C for 4 weeks. After thawing at room temperature for 2 h, the COT values were obtained from five sites on each cornea: central corneal thickness (CCOT), thickness at the 3, 6, 9, and 12 o'clock positions (3COT, 6COT, 9COT, and 12COT, respectively). For each device, measurements were performed three times at each designated site by one operator. Intraclass correlation coefficient, repeated‐measures analysis of variance, and Bland–Altman plot were performed for statistical analyses. Results: The mean CCOT was 839.0 ± 138.2, 857.6 ± 127.9, and 849.1 ± 132.8 μm for UP, UBM, and MM, respectively. There was no statistically significant difference among the measurements by all three devices ( P > 0.05). The statistical agreement among these devices was also shown to be good according to Bland–Altman plots. In addition, the values measured on the peripheral sites of the cornea (3COT, 6COT, 9COT, and 12COT) also revealed no significant differences among the three devices. The difference between CCOT and peripheral corneal thickness was also statistically insignificant. Conclusions: The UP, UBM, and MM measurements showed statistically similar COT values. The UBM and UP provided accurate and reliable measurements comparable to MM and could be used effectively for clinical COT measurements. … (more)
- Is Part Of:
- Veterinary ophthalmology. Volume 21:Number 4(2018)
- Journal:
- Veterinary ophthalmology
- Issue:
- Volume 21:Number 4(2018)
- Issue Display:
- Volume 21, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 21
- Issue:
- 4
- Issue Sort Value:
- 2018-0021-0004-0000
- Page Start:
- 339
- Page End:
- 346
- Publication Date:
- 2017-11-07
- Subjects:
- canine -- central corneal thickness -- frozen cornea -- peripheral corneal thickness -- ultrasound biomicroscopy -- ultrasound pachymetry
Veterinary ophthalmology -- Periodicals
636.08977 - Journal URLs:
- http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=vop ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1463-5224 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/vop.12509 ↗
- Languages:
- English
- ISSNs:
- 1463-5216
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9229.162000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6969.xml