Dose rate response of Digital Megavolt Imager detector for flattening filter‐free beams. (21st May 2018)
- Record Type:
- Journal Article
- Title:
- Dose rate response of Digital Megavolt Imager detector for flattening filter‐free beams. (21st May 2018)
- Main Title:
- Dose rate response of Digital Megavolt Imager detector for flattening filter‐free beams
- Authors:
- Xu, Zhigang
Kim, Jinkoo
Han, James
Hsia, An Ting
Ryu, Samuel - Abstract:
- Abstract: In this study we investigated the dose rate response characteristics of the Digital Megavolt Imager (DMI) detector, including panel saturation, linearity, and imager ghosting effects for flattening filter‐free (FFF) beams. The DMI detector dose rate response characteristics were measured as a function of dose rate on a Varian TrueBeam machine. Images were acquired at dose rates ranging from 400 to 1400 MU/min for 6XFFF and 400 to 2400 MU/min for 10XFFF. Line profiles and central portal doses derived from the images were analyzed and compared. The linearity was verified by acquiring images with incremental Monitor Unit (MU) ranging from 5 to 500 MU. Ghosting effects were studied at different dose rates. Finally, for validation, test plans with optimal fluence were created and measured with different dose rates. All test plans were analyzed with a Gamma criteria of 3%‐3 mm and 10% dose threshold. Our study showed that there was no panel saturation observed from the profile comparison even at the maximum dose rate of 2400 MU/min. The central portal doses showed a slight decrease (1.013–1.008 cGy/MU for 6XFFF, and 1.020–1.009 cGy/MU for 10XFFF) when dose rate increased (400–1400 MU/min for 6XFFF, and 400–2400 MU/min for 10XFFF). The linearity of the DMI detector response was better than 0.5% in the range of 20–500 MU for all energies. The residual image was extremely small and statistically undetectable. The Gamma index measured with the test plans decreased from 100%Abstract: In this study we investigated the dose rate response characteristics of the Digital Megavolt Imager (DMI) detector, including panel saturation, linearity, and imager ghosting effects for flattening filter‐free (FFF) beams. The DMI detector dose rate response characteristics were measured as a function of dose rate on a Varian TrueBeam machine. Images were acquired at dose rates ranging from 400 to 1400 MU/min for 6XFFF and 400 to 2400 MU/min for 10XFFF. Line profiles and central portal doses derived from the images were analyzed and compared. The linearity was verified by acquiring images with incremental Monitor Unit (MU) ranging from 5 to 500 MU. Ghosting effects were studied at different dose rates. Finally, for validation, test plans with optimal fluence were created and measured with different dose rates. All test plans were analyzed with a Gamma criteria of 3%‐3 mm and 10% dose threshold. Our study showed that there was no panel saturation observed from the profile comparison even at the maximum dose rate of 2400 MU/min. The central portal doses showed a slight decrease (1.013–1.008 cGy/MU for 6XFFF, and 1.020–1.009 cGy/MU for 10XFFF) when dose rate increased (400–1400 MU/min for 6XFFF, and 400–2400 MU/min for 10XFFF). The linearity of the DMI detector response was better than 0.5% in the range of 20–500 MU for all energies. The residual image was extremely small and statistically undetectable. The Gamma index measured with the test plans decreased from 100% to 97.8% for 6XFFF when dose rate increased from 400 to 1400 MU/min. For 10XFFF, the Gamma index decreased from 99.9% to 91.5% when dose rate increased from 400 to 2400 MU/min. We concluded that the Portal Dosimetry system for the TrueBeam using DMI detector can be reliably used for IMRT and VMAT QA for FFF energies. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 19:Number 4(2018)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 19:Number 4(2018)
- Issue Display:
- Volume 19, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 19
- Issue:
- 4
- Issue Sort Value:
- 2018-0019-0004-0000
- Page Start:
- 141
- Page End:
- 147
- Publication Date:
- 2018-05-21
- Subjects:
- FFF -- PDC -- IMRT -- VMAT -- EPID -- QA
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12358 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6882.xml