Multicenter clinical assessment of improved wearable multimodal convulsive seizure detectors. (4th October 2017)
- Record Type:
- Journal Article
- Title:
- Multicenter clinical assessment of improved wearable multimodal convulsive seizure detectors. (4th October 2017)
- Main Title:
- Multicenter clinical assessment of improved wearable multimodal convulsive seizure detectors
- Authors:
- Onorati, Francesco
Regalia, Giulia
Caborni, Chiara
Migliorini, Matteo
Bender, Daniel
Poh, Ming‐Zher
Frazier, Cherise
Kovitch Thropp, Eliana
Mynatt, Elizabeth D.
Bidwell, Jonathan
Mai, Roberto
LaFrance, W. Curt
Blum, Andrew S.
Friedman, Daniel
Loddenkemper, Tobias
Mohammadpour‐Touserkani, Fatemeh
Reinsberger, Claus
Tognetti, Simone
Picard, Rosalind W. - Abstract:
- Summary: Objective: New devices are needed for monitoring seizures, especially those associated with sudden unexpected death in epilepsy (SUDEP). They must be unobtrusive and automated, and provide false alarm rates (FARs) bearable in everyday life. This study quantifies the performance of new multimodal wrist‐worn convulsive seizure detectors. Methods: Hand‐annotated video‐electroencephalographic seizure events were collected from 69 patients at six clinical sites. Three different wristbands were used to record electrodermal activity (EDA) and accelerometer (ACM) signals, obtaining 5, 928 h of data, including 55 convulsive epileptic seizures (six focal tonic–clonic seizures and 49 focal to bilateral tonic–clonic seizures) from 22 patients. Recordings were analyzed offline to train and test two new machine learning classifiers and a published classifier based on EDA and ACM. Moreover, wristband data were analyzed to estimate seizure‐motion duration and autonomic responses. Results: The two novel classifiers consistently outperformed the previous detector. The most efficient (Classifier III) yielded sensitivity of 94.55%, and an FAR of 0.2 events/day. No nocturnal seizures were missed. Most patients had <1 false alarm every 4 days, with an FAR below their seizure frequency. When increasing the sensitivity to 100% (no missed seizures), the FAR is up to 13 times lower than with the previous detector. Furthermore, all detections occurred before the seizure ended, providingSummary: Objective: New devices are needed for monitoring seizures, especially those associated with sudden unexpected death in epilepsy (SUDEP). They must be unobtrusive and automated, and provide false alarm rates (FARs) bearable in everyday life. This study quantifies the performance of new multimodal wrist‐worn convulsive seizure detectors. Methods: Hand‐annotated video‐electroencephalographic seizure events were collected from 69 patients at six clinical sites. Three different wristbands were used to record electrodermal activity (EDA) and accelerometer (ACM) signals, obtaining 5, 928 h of data, including 55 convulsive epileptic seizures (six focal tonic–clonic seizures and 49 focal to bilateral tonic–clonic seizures) from 22 patients. Recordings were analyzed offline to train and test two new machine learning classifiers and a published classifier based on EDA and ACM. Moreover, wristband data were analyzed to estimate seizure‐motion duration and autonomic responses. Results: The two novel classifiers consistently outperformed the previous detector. The most efficient (Classifier III) yielded sensitivity of 94.55%, and an FAR of 0.2 events/day. No nocturnal seizures were missed. Most patients had <1 false alarm every 4 days, with an FAR below their seizure frequency. When increasing the sensitivity to 100% (no missed seizures), the FAR is up to 13 times lower than with the previous detector. Furthermore, all detections occurred before the seizure ended, providing reasonable latency (median = 29.3 s, range = 14.8–151 s). Automatically estimated seizure durations were correlated with true durations, enabling reliable annotations. Finally, EDA measurements confirmed the presence of postictal autonomic dysfunction, exhibiting a significant rise in 73% of the convulsive seizures. Significance: The proposed multimodal wrist‐worn convulsive seizure detectors provide seizure counts that are more accurate than previous automated detectors and typical patient self‐reports, while maintaining a tolerable FAR for ambulatory monitoring. Furthermore, the multimodal system provides an objective description of motor behavior and autonomic dysfunction, aimed at enriching seizure characterization, with potential utility for SUDEP warning. … (more)
- Is Part Of:
- Epilepsia. Volume 58:issue 11(2017)
- Journal:
- Epilepsia
- Issue:
- Volume 58:issue 11(2017)
- Issue Display:
- Volume 58, Issue 11 (2017)
- Year:
- 2017
- Volume:
- 58
- Issue:
- 11
- Issue Sort Value:
- 2017-0058-0011-0000
- Page Start:
- 1870
- Page End:
- 1879
- Publication Date:
- 2017-10-04
- Subjects:
- Epilepsy -- Convulsive seizures -- Electrodermal activity -- Machine learning
Epilepsy -- Periodicals
616.853 - Journal URLs:
- http://www.blackwell-synergy.com/servlet/useragent?func=showIssues&code=epi ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/epi.13899 ↗
- Languages:
- English
- ISSNs:
- 0013-9580
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3793.700000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6857.xml