Image quality and absorbed dose comparison of single‐ and dual‐source cone‐beam computed tomography. (17th April 2018)
- Record Type:
- Journal Article
- Title:
- Image quality and absorbed dose comparison of single‐ and dual‐source cone‐beam computed tomography. (17th April 2018)
- Main Title:
- Image quality and absorbed dose comparison of single‐ and dual‐source cone‐beam computed tomography
- Authors:
- Miura, Hideharu
Ozawa, Shuichi
Okazue, Toshiya
Kawakubo, Atsushi
Yamada, Kiyoshi
Nagata, Yasushi - Abstract:
- Abstract: Purpose: Dual‐source cone‐beam computed tomography (DCBCT) is currently available in the Vero4DRT image‐guided radiotherapy system. We evaluated the image quality and absorbed dose for DCBCT and compared the values with those for single‐source CBCT (SCBCT). Methods: Image uniformity, Hounsfield unit (HU) linearity, image contrast, and spatial resolution were evaluated using a Catphan phantom. The rotation angle for acquiring SCBCT and DCBCT images is 215° and 115°, respectively. The image uniformity was calculated using measurements obtained at the center and four peripheral positions. The HUs of seven materials inserted into the phantom were measured to evaluate HU linearity and image contrast. The Catphan phantom was scanned with a conventional CT scanner to measure the reference HU for each material. The spatial resolution was calculated using high‐resolution pattern modules. Image quality was analyzed using ImageJ software ver. 1.49. The absorbed dose was measured using a 0.6‐cm 3 ionization chamber with a 16‐cm‐diameter cylindrical phantom, at the center and four peripheral positions of the phantom, and calculated using weighted cone‐beam CT dose index (CBCTDIw ). Results: Compared with that of SCBCT, the image uniformity of DCBCT was slightly reduced. A strong linear correlation existed between the measured HU for DCBCT and the reference HU, although the linear regression slope was different from that of the reference HU. DCBCT had poorer image contrast thanAbstract: Purpose: Dual‐source cone‐beam computed tomography (DCBCT) is currently available in the Vero4DRT image‐guided radiotherapy system. We evaluated the image quality and absorbed dose for DCBCT and compared the values with those for single‐source CBCT (SCBCT). Methods: Image uniformity, Hounsfield unit (HU) linearity, image contrast, and spatial resolution were evaluated using a Catphan phantom. The rotation angle for acquiring SCBCT and DCBCT images is 215° and 115°, respectively. The image uniformity was calculated using measurements obtained at the center and four peripheral positions. The HUs of seven materials inserted into the phantom were measured to evaluate HU linearity and image contrast. The Catphan phantom was scanned with a conventional CT scanner to measure the reference HU for each material. The spatial resolution was calculated using high‐resolution pattern modules. Image quality was analyzed using ImageJ software ver. 1.49. The absorbed dose was measured using a 0.6‐cm 3 ionization chamber with a 16‐cm‐diameter cylindrical phantom, at the center and four peripheral positions of the phantom, and calculated using weighted cone‐beam CT dose index (CBCTDIw ). Results: Compared with that of SCBCT, the image uniformity of DCBCT was slightly reduced. A strong linear correlation existed between the measured HU for DCBCT and the reference HU, although the linear regression slope was different from that of the reference HU. DCBCT had poorer image contrast than did SCBCT, particularly with a high‐contrast material. There was no significant difference between the spatial resolutions of SCBCT and DCBCT. The absorbed dose for DCBCT was higher than that for SCBCT, because in DCBCT, the two x‐ray projections overlap between 45° and 70°. Conclusions: We found that the image quality was poorer and the absorbed dose was higher for DCBCT than for SCBCT in the Vero4DRT. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 19:Number 3(2018)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 19:Number 3(2018)
- Issue Display:
- Volume 19, Issue 3 (2018)
- Year:
- 2018
- Volume:
- 19
- Issue:
- 3
- Issue Sort Value:
- 2018-0019-0003-0000
- Page Start:
- 360
- Page End:
- 366
- Publication Date:
- 2018-04-17
- Subjects:
- absorbed dose -- dual‐source cone‐beam computed tomography -- image quality -- image‐guided radiation therapy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12328 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6800.xml