Optimisation of antithrombin resistance assay as a practical clinical laboratory test: Development of prothrombin activator using factors Xa/Va and automation of assay. (13th February 2018)
- Record Type:
- Journal Article
- Title:
- Optimisation of antithrombin resistance assay as a practical clinical laboratory test: Development of prothrombin activator using factors Xa/Va and automation of assay. (13th February 2018)
- Main Title:
- Optimisation of antithrombin resistance assay as a practical clinical laboratory test: Development of prothrombin activator using factors Xa/Va and automation of assay
- Authors:
- Tamura, S.
Suga, Y.
Tanamura, M.
Murata‐Kawakami, M.
Takagi, Y.
Hottori, Y.
Kakihara, M.
Suzuki, S.
Takagi, A.
Kojima, T. - Abstract:
- Abstract: Introduction: Antithrombin resistance (ATR) is a novel thrombotic risk in abnormal prothrombins. A manual ATR assay using Oxyuranus scutellatus (Ox) venom as a prothrombin activator was established for detecting antithrombin‐resistant prothrombin. However, this assay was limited because of Ox snake venom availability and its throughput capacity. Here, we have improved the ATR assay using bovine factors Xa and Va (FXa/Va) as prothrombin activators and have optimised assay conditions for an automated instrument (ACL TOP 500). Methods: Diluted plasma was incubated with a prothrombin activator mix (phospholipids, CaCl2, and bovine FXa/Va), followed by inactivation with antithrombin for 10, 20 and 30 minutes. We added a chromogenic substrate S‐2238, and assessed changes in absorbance/min at 405 nm. We also adapted assay conditions for ACL TOP 500. Results: Optimum conditions for FXa/Va treatment were 6.25% phospholipids, 5 mM CaCL2, 0.01 μg/mL FXa and 0.1 μg/mL FVa. ATR assay kinetics with the FXa/Va activator was comparable with that with the Ox activator in heterozygous reconstituted plasma with the recombinant wild‐type or antithrombin‐resistant prothrombin. Using ACL TOP 500, optimum conditions for the FXa/Va treatment were 10.0% phospholipids, 5 mM CaCl2, 0.02 μg/mL FXa and 0.2 μg/mL FVa. The automated ATR assay with the FXa/Va activator demonstrated good detectability for antithrombin‐resistant prothrombin in plasma from a heterozygous carrier with prothrombinAbstract: Introduction: Antithrombin resistance (ATR) is a novel thrombotic risk in abnormal prothrombins. A manual ATR assay using Oxyuranus scutellatus (Ox) venom as a prothrombin activator was established for detecting antithrombin‐resistant prothrombin. However, this assay was limited because of Ox snake venom availability and its throughput capacity. Here, we have improved the ATR assay using bovine factors Xa and Va (FXa/Va) as prothrombin activators and have optimised assay conditions for an automated instrument (ACL TOP 500). Methods: Diluted plasma was incubated with a prothrombin activator mix (phospholipids, CaCl2, and bovine FXa/Va), followed by inactivation with antithrombin for 10, 20 and 30 minutes. We added a chromogenic substrate S‐2238, and assessed changes in absorbance/min at 405 nm. We also adapted assay conditions for ACL TOP 500. Results: Optimum conditions for FXa/Va treatment were 6.25% phospholipids, 5 mM CaCL2, 0.01 μg/mL FXa and 0.1 μg/mL FVa. ATR assay kinetics with the FXa/Va activator was comparable with that with the Ox activator in heterozygous reconstituted plasma with the recombinant wild‐type or antithrombin‐resistant prothrombin. Using ACL TOP 500, optimum conditions for the FXa/Va treatment were 10.0% phospholipids, 5 mM CaCl2, 0.02 μg/mL FXa and 0.2 μg/mL FVa. The automated ATR assay with the FXa/Va activator demonstrated good detectability for antithrombin‐resistant prothrombin in plasma from a heterozygous carrier with prothrombin Yukuhashi or Belgrade. Conclusion: We optimised the ATR assay with the FXa/Va activator and adapted the assay for ACL TOP 500; the assay showed the ability to clearly detect antithrombin‐resistant prothrombin in manual and automated procedures. … (more)
- Is Part Of:
- International journal of laboratory hematology. Volume 40:Number 3(2018:Jun.)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 40:Number 3(2018:Jun.)
- Issue Display:
- Volume 40, Issue 3 (2018)
- Year:
- 2018
- Volume:
- 40
- Issue:
- 3
- Issue Sort Value:
- 2018-0040-0003-0000
- Page Start:
- 312
- Page End:
- 319
- Publication Date:
- 2018-02-13
- Subjects:
- ACL TOP 500 -- antithrombin resistance -- FXa/Va activator -- Ox venom -- prothrombin
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.12786 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 6770.xml