Biomechanical optimization of subject-specific implant positioning for femoral head resurfacing to reduce fracture risk. (July 2016)
- Record Type:
- Journal Article
- Title:
- Biomechanical optimization of subject-specific implant positioning for femoral head resurfacing to reduce fracture risk. (July 2016)
- Main Title:
- Biomechanical optimization of subject-specific implant positioning for femoral head resurfacing to reduce fracture risk
- Authors:
- Miles, Brad
Kolos, Elizabeth
Appleyard, Richard
Theodore, Willy
Zheng, Keke
Li, Qing
Ruys, Andrew J - Abstract:
- Peri-prosthetic femoral neck fracture after femoral head resurfacing can be either patient-related or surgical technique-related. The study aimed to develop a patient-specific finite element modelling technique that can reliably predict an optimal implant position and give minimal strain in the peri-prosthetic bone tissue, thereby reducing the risk of peri-prosthetic femoral neck fracture. The subject-specific finite element modelling was integrated with optimization techniques including design of experiments to best possibly position the implant for achieving minimal strain for femoral head resurfacing. Sample space was defined by varying the floating point to find the extremes at which the cylindrical reaming operation actually cuts into the femoral neck causing a notch during hip resurfacing surgery. The study showed that the location of the maximum strain, for all non-notching positions, was on the superior femoral neck, in the peri-prosthetic bone tissue. It demonstrated that varus positioning resulted in a higher strain, while valgus positioning reduced the strain, and further that neutral version had a lower strain.
- Is Part Of:
- Proceedings of the Institution of Mechanical Engineers. Volume 230:Number 7(2016:Jul.)
- Journal:
- Proceedings of the Institution of Mechanical Engineers
- Issue:
- Volume 230:Number 7(2016:Jul.)
- Issue Display:
- Volume 230, Issue 7 (2016)
- Year:
- 2016
- Volume:
- 230
- Issue:
- 7
- Issue Sort Value:
- 2016-0230-0007-0000
- Page Start:
- 668
- Page End:
- 674
- Publication Date:
- 2016-07
- Subjects:
- Femoral head resurfacing -- finite element analysis -- goal-driven optimization -- design of experiments
Biomedical engineering -- Periodicals
Medical instruments and apparatus -- Periodicals
610.28 - Journal URLs:
- http://pih.sagepub.com/ ↗
http://journals.pepublishing.com/content/119779 ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0954411916644633 ↗
- Languages:
- English
- ISSNs:
- 0954-4119
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6676.xml