Direct 4D PET MLEM reconstruction of parametric images using the simplified reference tissue model with the basis function method for [11C]raclopride. (20th May 2015)
- Record Type:
- Journal Article
- Title:
- Direct 4D PET MLEM reconstruction of parametric images using the simplified reference tissue model with the basis function method for [11C]raclopride. (20th May 2015)
- Main Title:
- Direct 4D PET MLEM reconstruction of parametric images using the simplified reference tissue model with the basis function method for [11C]raclopride
- Authors:
- Gravel, Paul
Reader, Andrew J - Abstract:
- Abstract: This work assesses the one-step late maximum likelihood expectation maximization (OSL-MLEM) 4D PET reconstruction algorithm for direct estimation of parametric images from raw PET data when using the simplified reference tissue model with the basis function method (SRTM-BFM) for the kinetic analysis. To date, the OSL-MLEM method has been evaluated using kinetic models based on two-tissue compartments with an irreversible component. We extend the evaluation of this method for two-tissue compartments with a reversible component, using SRTM-BFM on simulated 3D + time data sets (with use of [ 11 C]raclopride time-activity curves from real data) and on real data sets acquired with the high resolution research tomograph. The performance of the proposed method is evaluated by comparing voxel-level binding potential ( BP ND ) estimates with those obtained from conventional post-reconstruction kinetic parameter estimation. For the commonly chosen number of iterations used in practice, our results show that for the 3D + time simulation, the direct method delivers results with lower % RMSE at the normal count level (decreases of 9–10 percentage points, corresponding to a 38–44% reduction), and also at low count levels (decreases of 17–21 percentage points, corresponding to a 26–36% reduction). As for the real 3D data set, the results obtained follow a similar trend, with the direct reconstruction method offering a 21% decrease in % CV compared to the post reconstructionAbstract: This work assesses the one-step late maximum likelihood expectation maximization (OSL-MLEM) 4D PET reconstruction algorithm for direct estimation of parametric images from raw PET data when using the simplified reference tissue model with the basis function method (SRTM-BFM) for the kinetic analysis. To date, the OSL-MLEM method has been evaluated using kinetic models based on two-tissue compartments with an irreversible component. We extend the evaluation of this method for two-tissue compartments with a reversible component, using SRTM-BFM on simulated 3D + time data sets (with use of [ 11 C]raclopride time-activity curves from real data) and on real data sets acquired with the high resolution research tomograph. The performance of the proposed method is evaluated by comparing voxel-level binding potential ( BP ND ) estimates with those obtained from conventional post-reconstruction kinetic parameter estimation. For the commonly chosen number of iterations used in practice, our results show that for the 3D + time simulation, the direct method delivers results with lower % RMSE at the normal count level (decreases of 9–10 percentage points, corresponding to a 38–44% reduction), and also at low count levels (decreases of 17–21 percentage points, corresponding to a 26–36% reduction). As for the real 3D data set, the results obtained follow a similar trend, with the direct reconstruction method offering a 21% decrease in % CV compared to the post reconstruction method at low count levels. Thus, based on the results presented herein, using the SRTM-BFM kinetic model in conjunction with the OSL-MLEM direct 4D PET MLEM reconstruction method offers an improvement in performance when compared to conventional post reconstruction methods. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 60:Number 11(2015:Jun.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 60:Number 11(2015:Jun.)
- Issue Display:
- Volume 60, Issue 11 (2015)
- Year:
- 2015
- Volume:
- 60
- Issue:
- 11
- Issue Sort Value:
- 2015-0060-0011-0000
- Page Start:
- 4533
- Page End:
- 4549
- Publication Date:
- 2015-05-20
- Subjects:
- positron emission tomography -- image reconstruction -- tracer kinetic modeling -- binding potential
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/0031-9155/60/11/4533 ↗