Adaptive iterative dose reduction 3D (AIDR 3D) vs. filtered back projection: radiation dose reduction capabilities of wide volume and helical scanning techniques on area-detector CT in a chest phantom study. (June 2016)
- Record Type:
- Journal Article
- Title:
- Adaptive iterative dose reduction 3D (AIDR 3D) vs. filtered back projection: radiation dose reduction capabilities of wide volume and helical scanning techniques on area-detector CT in a chest phantom study. (June 2016)
- Main Title:
- Adaptive iterative dose reduction 3D (AIDR 3D) vs. filtered back projection: radiation dose reduction capabilities of wide volume and helical scanning techniques on area-detector CT in a chest phantom study
- Authors:
- Seki, Shinichiro
Koyama, Hisanobu
Ohno, Yoshiharu
Matsumoto, Sumiaki
Inokawa, Hiroyasu
Sugihara, Naoki
Sugimura, Kazuro - Abstract:
- Background: Computed tomography (CT) has important roles for lung cancer screening, and therefore radiation dose reduction by using iterative reconstruction technique and scanning methods receive widespread attention. Purpose: To evaluate the effect of two reconstruction techniques (filtered back projection [FBP] and adaptive iterative dose reduction using three-dimensional processing [AIDR 3D]) and two acquisition techniques (wide-volume scan [WVS] and helical scan as 64-detector-row CT [64HS]) on the lung nodule identifications of using a chest phantom. Material and Methods: A chest CT phantom including lung nodules was scanned using WVS and 64HS at nine different tube currents (TCs; range, 270–10 mA). All CT datasets were reconstructed with AIDR 3D and FBP. Standard deviation (SD) measurements by region of interest placement and qualitative nodule identifications were statistically compared. 64HS and WVS were evaluated separately, and FBP images acquired with 270 mA was defined as the standard reference. Results: SDs of all datasets with AIDR 3D showed no significant differences ( P > 0.05) with standard reference. When comparing nodule identifications, area under the curve on WVS with AIDR 3D with TC <30 mA, on 64HS with AIDR 3D with TC <40 mA, and on reconstructions with FBP and each scan method with TC <60 mA was significantly lower than with standard reference ( P < 0.05). With the same TC and reconstruction, SDs and nodule identifications of WVS were notBackground: Computed tomography (CT) has important roles for lung cancer screening, and therefore radiation dose reduction by using iterative reconstruction technique and scanning methods receive widespread attention. Purpose: To evaluate the effect of two reconstruction techniques (filtered back projection [FBP] and adaptive iterative dose reduction using three-dimensional processing [AIDR 3D]) and two acquisition techniques (wide-volume scan [WVS] and helical scan as 64-detector-row CT [64HS]) on the lung nodule identifications of using a chest phantom. Material and Methods: A chest CT phantom including lung nodules was scanned using WVS and 64HS at nine different tube currents (TCs; range, 270–10 mA). All CT datasets were reconstructed with AIDR 3D and FBP. Standard deviation (SD) measurements by region of interest placement and qualitative nodule identifications were statistically compared. 64HS and WVS were evaluated separately, and FBP images acquired with 270 mA was defined as the standard reference. Results: SDs of all datasets with AIDR 3D showed no significant differences ( P > 0.05) with standard reference. When comparing nodule identifications, area under the curve on WVS with AIDR 3D with TC <30 mA, on 64HS with AIDR 3D with TC <40 mA, and on reconstructions with FBP and each scan method with TC <60 mA was significantly lower than with standard reference ( P < 0.05). With the same TC and reconstruction, SDs and nodule identifications of WVS were not significantly different from 64HS ( P > 0.05). Conclusion: In term of SD of lung parenchyma and nodule identification, AIDR 3D can achieve more radiation dose reduction than FBP and there is no significant different between WVS and 64HS. … (more)
- Is Part Of:
- Acta radiologica. Volume 57:Number 6(2016:Jun.)
- Journal:
- Acta radiologica
- Issue:
- Volume 57:Number 6(2016:Jun.)
- Issue Display:
- Volume 57, Issue 6 (2016)
- Year:
- 2016
- Volume:
- 57
- Issue:
- 6
- Issue Sort Value:
- 2016-0057-0006-0000
- Page Start:
- 684
- Page End:
- 690
- Publication Date:
- 2016-06
- Subjects:
- Lung -- computed tomography (CT) -- radiation dose -- iterative reconstruction technique
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1177/0284185115603418 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6599.xml