Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study. (February 2016)
- Record Type:
- Journal Article
- Title:
- Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study. (February 2016)
- Main Title:
- Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film
- Authors:
- Steenbeke, Femke
Gevaert, Thierry
Tournel, Koen
Engels, Benedikt
Verellen, Dirk
Storme, Guy
De Ridder, Mark - Abstract:
- Purpose: Radiochromic EBT3 film is gaining acceptance as a valuable dosimetry system for high-energy photon beams. The advantages of these films over other dosimetry systems are low spectral sensitivity and high spatial resolution. The aim of this study was to validate EBT3 film as a dosimeter for machine and treatment quality assurance (QA) of a 50-kV radiotherapy unit. Methods and Materials: Absolute and relative doses were acquired using EBT3 GafChromic films and compared to a parallel-plate ionization chamber (IC), the standard IC for low-energy X-rays. EBT3 was also used to evaluate beam profiles and output factors. Two films above each other, mimicking the clinical situation of a dosimeter on top of the skin, were simultaneously irradiated to evaluate EBT3 as in vivo dosimeter. All films were irradiated for 3 minutes, which corresponds with a surface dose of 3.25 ± 0.07 Gy. Results: A fifth-order polynomial function was found to be the best fit for the calibration curves. Good agreement between IC and EBT3 was found for absolute (0.92% for green and red color channels) and relative (1.2% and 1.0% for green and red color channels, respectively) dosimetry. Output factors for IC and EBT3 were comparable within 2.04% and 1.02% for the green and red color channels, respectively. Flatness and symmetry at the surface were within 2%. By applying film as in vivo dosimeter, an absorption of 4.70% needs to be taken into account with respect to the surface dose. Conclusion: EBT3Purpose: Radiochromic EBT3 film is gaining acceptance as a valuable dosimetry system for high-energy photon beams. The advantages of these films over other dosimetry systems are low spectral sensitivity and high spatial resolution. The aim of this study was to validate EBT3 film as a dosimeter for machine and treatment quality assurance (QA) of a 50-kV radiotherapy unit. Methods and Materials: Absolute and relative doses were acquired using EBT3 GafChromic films and compared to a parallel-plate ionization chamber (IC), the standard IC for low-energy X-rays. EBT3 was also used to evaluate beam profiles and output factors. Two films above each other, mimicking the clinical situation of a dosimeter on top of the skin, were simultaneously irradiated to evaluate EBT3 as in vivo dosimeter. All films were irradiated for 3 minutes, which corresponds with a surface dose of 3.25 ± 0.07 Gy. Results: A fifth-order polynomial function was found to be the best fit for the calibration curves. Good agreement between IC and EBT3 was found for absolute (0.92% for green and red color channels) and relative (1.2% and 1.0% for green and red color channels, respectively) dosimetry. Output factors for IC and EBT3 were comparable within 2.04% and 1.02% for the green and red color channels, respectively. Flatness and symmetry at the surface were within 2%. By applying film as in vivo dosimeter, an absorption of 4.70% needs to be taken into account with respect to the surface dose. Conclusion: EBT3 GafChromic film is a feasible and valuable QA and dosimetry tool for a 50-kV radiotherapy unit. EBT3 can be used for absolute and relative dosimetry, measurement of output factors and beam profiles. In vivo patient-specific QA can also be performed if one corrects for the dose absorption of the film. … (more)
- Is Part Of:
- Technology in cancer research & treatment. Volume 15:Number 1(2016:Feb.)
- Journal:
- Technology in cancer research & treatment
- Issue:
- Volume 15:Number 1(2016:Feb.)
- Issue Display:
- Volume 15, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 15
- Issue:
- 1
- Issue Sort Value:
- 2016-0015-0001-0000
- Page Start:
- 163
- Page End:
- 170
- Publication Date:
- 2016-02
- Subjects:
- contact X-ray radiotherapy -- low-energy dosimetry -- EBT3 GafChromic film -- in vivo dosimetry -- absolute and relative dosimetry
Oncology -- Periodicals
Cancer -- Diagnosis -- Periodicals
Cancer -- Treatment -- Technological innovations -- Periodicals
616.994 - Journal URLs:
- http://tct.sagepub.com/ ↗
http://www.tcrt.org ↗
http://www.sagepub.com ↗ - DOI:
- 10.1177/1533034614565910 ↗
- Languages:
- English
- ISSNs:
- 1533-0346
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6581.xml