Development of a dual-energy silicon X-ray diode and its application to gadolinium imaging. (June 2015)
- Record Type:
- Journal Article
- Title:
- Development of a dual-energy silicon X-ray diode and its application to gadolinium imaging. (June 2015)
- Main Title:
- Development of a dual-energy silicon X-ray diode and its application to gadolinium imaging
- Authors:
- Sato, Yuichi
Sato, Eiichi
Ehara, Shigeru
Oda, Yasuyuki
Hagiwara, Osahiko
Matsukiyo, Hiroshi
Enomoto, Toshiyuki
Watanabe, Manabu
Kusachi, Shinya - Abstract:
- Abstract: To perform dual-energy X-ray imaging, we developed a dual-energy silicon X-ray diode (DE-Si-XD) consisting of two ceramic-substrate silicon X-ray diodes (Si-XD) and a 0.2-mm-thick copper filter. The Si-XD is a high-sensitivity Si photodiode selected for detecting X-rays. In the front Si-XD, X-ray photons from an X-ray tube are directly detected. Because low-energy photons are absorbed by the front Si-XD and the filter, the average photon energy increases when the back Si-XD is used. In the front Si-XD, the photocurrents flowing through the Si-XD are converted into voltages and amplified using current–voltage and voltage–voltage (V–V) amplifiers. The output from the V–V amplifier is input to an analog-digital converter through an integrator for smoothing the voltage. The same amplification method is also used in the back Si-XD. Dual-energy computed tomography (DE–CT) is accomplished by repeated linear scans and rotations of the object, and two projection curves of the object are obtained simultaneously by linear scanning at a tube voltage of 90 kV and a current of 1.0 mA. In the DE–CT, the exposure time for obtaining a tomogram is 10 min with scan steps of 0.5 mm and rotation steps of 1.0°. Using gadolinium-based contrast media, energy subtraction was performed. Highlights: Dual-energy X-ray diode consists of two Si diodes and a Cu filter. Low and high-energy X-rays are detected using front and back diodes. Two-different-energy tomograms were easily obtainedAbstract: To perform dual-energy X-ray imaging, we developed a dual-energy silicon X-ray diode (DE-Si-XD) consisting of two ceramic-substrate silicon X-ray diodes (Si-XD) and a 0.2-mm-thick copper filter. The Si-XD is a high-sensitivity Si photodiode selected for detecting X-rays. In the front Si-XD, X-ray photons from an X-ray tube are directly detected. Because low-energy photons are absorbed by the front Si-XD and the filter, the average photon energy increases when the back Si-XD is used. In the front Si-XD, the photocurrents flowing through the Si-XD are converted into voltages and amplified using current–voltage and voltage–voltage (V–V) amplifiers. The output from the V–V amplifier is input to an analog-digital converter through an integrator for smoothing the voltage. The same amplification method is also used in the back Si-XD. Dual-energy computed tomography (DE–CT) is accomplished by repeated linear scans and rotations of the object, and two projection curves of the object are obtained simultaneously by linear scanning at a tube voltage of 90 kV and a current of 1.0 mA. In the DE–CT, the exposure time for obtaining a tomogram is 10 min with scan steps of 0.5 mm and rotation steps of 1.0°. Using gadolinium-based contrast media, energy subtraction was performed. Highlights: Dual-energy X-ray diode consists of two Si diodes and a Cu filter. Low and high-energy X-rays are detected using front and back diodes. Two-different-energy tomograms were easily obtained simultaneously. Gd-K-edge CT was accomplished using the back diode. Energy subtraction was performed easily to image a target object. … (more)
- Is Part Of:
- Radiation measurements. Volume 77(2015:Jun.)
- Journal:
- Radiation measurements
- Issue:
- Volume 77(2015:Jun.)
- Issue Display:
- Volume 77 (2015)
- Year:
- 2015
- Volume:
- 77
- Issue Sort Value:
- 2015-0077-0000-0000
- Page Start:
- 12
- Page End:
- 17
- Publication Date:
- 2015-06
- Subjects:
- Dual-energy X-ray diode -- Double Si diode -- Dual-energy CT -- Gd-K-edge CT -- Energy subtraction
Nuclear emulsions -- Periodicals
Particle tracks (Nuclear physics) -- Periodicals
Thermoluminescence -- Periodicals
Cosmic rays -- Periodicals
Radiation -- Measurement -- Periodicals
Radiometry -- Periodicals
Radiation Monitoring -- Periodicals
Émulsions nucléaires -- Périodiques
Particules (Physique nucléaire) -- Traces -- Périodiques
Thermoluminescence -- Périodiques
Rayonnement cosmique -- Périodiques
Radiométrie -- Périodiques
539.77 - Journal URLs:
- http://www.sciencedirect.com/science/journal/13504487 ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiation-measurements/ ↗ - DOI:
- 10.1016/j.radmeas.2015.04.003 ↗
- Languages:
- English
- ISSNs:
- 1350-4487
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7227.973000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6573.xml