A Multicenter Performance Evaluation of a Blood Glucose Monitoring System in 21 Leading Hospitals in Spain. (January 2016)
- Record Type:
- Journal Article
- Title:
- A Multicenter Performance Evaluation of a Blood Glucose Monitoring System in 21 Leading Hospitals in Spain. (January 2016)
- Main Title:
- A Multicenter Performance Evaluation of a Blood Glucose Monitoring System in 21 Leading Hospitals in Spain
- Authors:
- Bedini, José Luis
Wallace, Jane F.
Petruschke, Thorsten
Pardo, Scott - Abstract:
- Background: Self-monitoring of blood glucose is crucial for the effective self-management of diabetes. The present study evaluated the accuracy of the Contour® XT blood glucose monitoring system (BGMS) compared to the reference method in a large multicenter study under routine lab conditions at each hospital site. Methods: This study was conducted at 21 leading hospitals in Spain using leftover whole blood samples (n = 2100). Samples were tested with the BGMS using 1 commercial strip lot and the local laboratory hexokinase method. BGMS accuracy was assessed and results were compared to ISO 15197:2013 accuracy limit criteria and by using mean absolute relative difference analysis (MARD), consensus (Parkes) error grid (CEG), and surveillance error grid analyses (SEG). Results: Pooled analysis of 2100 measurements from all sites showed that 99.43% of the BGMS results were within the ranges accepted by the accuracy limit criteria. The overall MARD was 3.85%. MARD was 4.47% for glucose concentrations < 70 mg/dL and 3.81% for concentrations of 70-300 mg/dL. In CEG, most results (99.8%) were within zone A ("no effect on clinical action"); the remaining ones (0.2%) were in zone B ("little to no effect on clinical action"). The SEG analysis showed that most of the results (98.4%) were in the "no risk" zone, with the remaining results in the "slight, lower" risk zone. Conclusions: This is the largest multicenter study of Contour XT BGMS to date, and shows that this BGMS meets the ISOBackground: Self-monitoring of blood glucose is crucial for the effective self-management of diabetes. The present study evaluated the accuracy of the Contour® XT blood glucose monitoring system (BGMS) compared to the reference method in a large multicenter study under routine lab conditions at each hospital site. Methods: This study was conducted at 21 leading hospitals in Spain using leftover whole blood samples (n = 2100). Samples were tested with the BGMS using 1 commercial strip lot and the local laboratory hexokinase method. BGMS accuracy was assessed and results were compared to ISO 15197:2013 accuracy limit criteria and by using mean absolute relative difference analysis (MARD), consensus (Parkes) error grid (CEG), and surveillance error grid analyses (SEG). Results: Pooled analysis of 2100 measurements from all sites showed that 99.43% of the BGMS results were within the ranges accepted by the accuracy limit criteria. The overall MARD was 3.85%. MARD was 4.47% for glucose concentrations < 70 mg/dL and 3.81% for concentrations of 70-300 mg/dL. In CEG, most results (99.8%) were within zone A ("no effect on clinical action"); the remaining ones (0.2%) were in zone B ("little to no effect on clinical action"). The SEG analysis showed that most of the results (98.4%) were in the "no risk" zone, with the remaining results in the "slight, lower" risk zone. Conclusions: This is the largest multicenter study of Contour XT BGMS to date, and shows that this BGMS meets the ISO 15197:2013 accuracy limit criteria under local routine conditions in 21 leading Spanish hospitals. … (more)
- Is Part Of:
- Journal of diabetes science and technology. Volume 10:Number 1(2016:Jan.)
- Journal:
- Journal of diabetes science and technology
- Issue:
- Volume 10:Number 1(2016:Jan.)
- Issue Display:
- Volume 10, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 10
- Issue:
- 1
- Issue Sort Value:
- 2016-0010-0001-0000
- Page Start:
- 93
- Page End:
- 100
- Publication Date:
- 2016-01
- Subjects:
- accuracy -- blood glucose monitoring system -- diabetes mellitus -- consensus (Parkes) error grid -- surveillance error grid -- Contour XT
Diabetes -- Periodicals
Medical technology -- Periodicals
Diabetes Mellitus -- Periodicals
616.462005 - Journal URLs:
- http://ejournals.ebsco.com/direct.asp?JournalID=712321 ↗
http://www.jodsat.org/about.html ↗
http://online.sagepub.com/ ↗ - DOI:
- 10.1177/1932296815598777 ↗