A radiographic comparison of human airway anatomy and airway manikins – Implications for manikin-based testing of artificial airways. (July 2015)
- Record Type:
- Journal Article
- Title:
- A radiographic comparison of human airway anatomy and airway manikins – Implications for manikin-based testing of artificial airways. (July 2015)
- Main Title:
- A radiographic comparison of human airway anatomy and airway manikins – Implications for manikin-based testing of artificial airways
- Authors:
- Schalk, Richard
Eichler, Kathrin
Bergold, Martin N.
Weber, Christian F.
Zacharowski, Kai
Meininger, Dirk
Byhahn, Christian
Mutlak, Haitham - Abstract:
- Abstract: Objective: The aim of this prospective, single-center, observational study was to investigate the accuracy of modeling and reproduction of human anatomical dimensions in manikins by comparing radiographic upper airway measurements of 13 different models with humans. Methods: 13 commonly used airway manikins (male or female anatomy based) and 47 controls (adult humans, 37 male, 10 female) were investigated using a mediosagittal and axial cervical spine CT scan. For anatomical comparison six human upper airway target structures, the following were measured: Oblique diameter of the tongue through the center, horizontal distance between the center point of the tongue and the posterior pharyngeal wall, horizontal distance between the vallecula and the posterior pharyngeal wall, distance of the upper oesophageal orifice length of epiglottis distance at the narrowest part of the trachea. Furthermore, the cross-section of the trachea in axial view and the cross-section of the upper oesophageal orifice in the same section was calculated. All measurements were compared gender specific, if the gender was non-specified with the whole sample. Results: None of the included 13 different airway manikins matched anatomy in human controls ( n = 47) in all of the six measurements. The Laerdal Airway Management Trainer, however, replicated human airway anatomy at least satisfactorily. Conclusion: This investigation showed that all of the examined manikins did not replicate humanAbstract: Objective: The aim of this prospective, single-center, observational study was to investigate the accuracy of modeling and reproduction of human anatomical dimensions in manikins by comparing radiographic upper airway measurements of 13 different models with humans. Methods: 13 commonly used airway manikins (male or female anatomy based) and 47 controls (adult humans, 37 male, 10 female) were investigated using a mediosagittal and axial cervical spine CT scan. For anatomical comparison six human upper airway target structures, the following were measured: Oblique diameter of the tongue through the center, horizontal distance between the center point of the tongue and the posterior pharyngeal wall, horizontal distance between the vallecula and the posterior pharyngeal wall, distance of the upper oesophageal orifice length of epiglottis distance at the narrowest part of the trachea. Furthermore, the cross-section of the trachea in axial view and the cross-section of the upper oesophageal orifice in the same section was calculated. All measurements were compared gender specific, if the gender was non-specified with the whole sample. Results: None of the included 13 different airway manikins matched anatomy in human controls ( n = 47) in all of the six measurements. The Laerdal Airway Management Trainer, however, replicated human airway anatomy at least satisfactorily. Conclusion: This investigation showed that all of the examined manikins did not replicate human anatomy. Manikins should therefore be selected cautiously, depending on the type of airway securing procedure. Their widespread use as a replacement for in vivo trials in the field of airway management needs to be reconsidered. … (more)
- Is Part Of:
- Resuscitation. Volume 92(2015)
- Journal:
- Resuscitation
- Issue:
- Volume 92(2015)
- Issue Display:
- Volume 92, Issue 2015 (2015)
- Year:
- 2015
- Volume:
- 92
- Issue:
- 2015
- Issue Sort Value:
- 2015-0092-2015-0000
- Page Start:
- 129
- Page End:
- 136
- Publication Date:
- 2015-07
- Subjects:
- Airway management -- Manikin -- Intratracheal intubation -- Anatomic models -- Patient simulation
Resuscitation -- Periodicals
Resuscitation -- Periodicals
Réanimation -- Périodiques
Electronic journals
616.025 - Journal URLs:
- http://www.sciencedirect.com/science/journal/03009572 ↗
http://www.resuscitationjournal.com/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/03009572 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/03009572 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.resuscitation.2015.05.001 ↗
- Languages:
- English
- ISSNs:
- 0300-9572
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7785.420000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6527.xml