Characterization of Tissue Response to Impact Loads Delivered Using a Hand-Held Instrument for Studying Articular Cartilage Injury. (October 2015)
- Record Type:
- Journal Article
- Title:
- Characterization of Tissue Response to Impact Loads Delivered Using a Hand-Held Instrument for Studying Articular Cartilage Injury. (October 2015)
- Main Title:
- Characterization of Tissue Response to Impact Loads Delivered Using a Hand-Held Instrument for Studying Articular Cartilage Injury
- Authors:
- Bonnevie, Edward D.
Delco, Michelle L.
Fortier, Lisa A.
Alexander, Peter G.
Tuan, Rocky S.
Bonassar, Lawrence J. - Abstract:
- Objective: The objective of this study was to fully characterize the mechanics of an in vivo impactor and correlate the mechanics with superficial cracking of articular surfaces. Design: A spring-loaded impactor was used to apply energy-controlled impacts to the articular surfaces of neonatal bovine cartilage. The simultaneous use of a load cell and displacement sensor provided measurements of stress, stress rate, strain, strain rate, and strain energy density. Application of India ink after impact was used to correlate the mechanical inputs during impact with the resulting severity of tissue damage. Additionally, a signal processing method to deconvolve inertial stresses from impact stresses was developed and validated. Results: Impact models fit the data well (root mean square error average ~0.09) and provided a fully characterized impact. Correlation analysis between mechanical inputs and degree of superficial cracking made visible through India ink application provided significant positive correlations for stress and stress rate with degree of surface cracking ( R 2 = 0.7398 and R 2 = 0.5262, respectively). Ranges of impact parameters were 7 to 21 MPa, 6 to 40 GPa/s, 0.16 to 0.38, 87 to 236 s −1, and 0.3 to 1.1 MJ/m 3 for stress, stress rate, strain, strain rate, and strain energy density, respectively. Thresholds for damage for all inputs were determined at 13 MPa, 15 GPa/s, 0.23, 160 s −1, and 0.59 MJ/m 3 for this system. Conclusions: This study provided the mechanicalObjective: The objective of this study was to fully characterize the mechanics of an in vivo impactor and correlate the mechanics with superficial cracking of articular surfaces. Design: A spring-loaded impactor was used to apply energy-controlled impacts to the articular surfaces of neonatal bovine cartilage. The simultaneous use of a load cell and displacement sensor provided measurements of stress, stress rate, strain, strain rate, and strain energy density. Application of India ink after impact was used to correlate the mechanical inputs during impact with the resulting severity of tissue damage. Additionally, a signal processing method to deconvolve inertial stresses from impact stresses was developed and validated. Results: Impact models fit the data well (root mean square error average ~0.09) and provided a fully characterized impact. Correlation analysis between mechanical inputs and degree of superficial cracking made visible through India ink application provided significant positive correlations for stress and stress rate with degree of surface cracking ( R 2 = 0.7398 and R 2 = 0.5262, respectively). Ranges of impact parameters were 7 to 21 MPa, 6 to 40 GPa/s, 0.16 to 0.38, 87 to 236 s −1, and 0.3 to 1.1 MJ/m 3 for stress, stress rate, strain, strain rate, and strain energy density, respectively. Thresholds for damage for all inputs were determined at 13 MPa, 15 GPa/s, 0.23, 160 s −1, and 0.59 MJ/m 3 for this system. Conclusions: This study provided the mechanical basis for use of a portable, sterilizable, and maneuverable impacting device. Use of this device enables controlled impact loads in vitro or in vivo to connect mechanistic studies with long-term monitoring of disease progression. … (more)
- Is Part Of:
- Cartilage. Volume 6:Number 4(2015)
- Journal:
- Cartilage
- Issue:
- Volume 6:Number 4(2015)
- Issue Display:
- Volume 6, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 6
- Issue:
- 4
- Issue Sort Value:
- 2015-0006-0004-0000
- Page Start:
- 226
- Page End:
- 232
- Publication Date:
- 2015-10
- Subjects:
- articular cartilage -- in vitro -- posttraumatic arthritis
Cartilage -- Periodicals
616.7 - Journal URLs:
- http://car.sagepub.com/ ↗
http://www.uk.sagepub.com ↗ - DOI:
- 10.1177/1947603515595071 ↗
- Languages:
- English
- ISSNs:
- 1947-6035
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6507.xml