Clinically oriented device programming in bradycardia patients: part 2 (atrioventricular blocks and neurally mediated syncope). Proposals from AIAC (Italian Association of Arrhythmology and Cardiac Pacing). Issue 4 (April 2018)
- Record Type:
- Journal Article
- Title:
- Clinically oriented device programming in bradycardia patients: part 2 (atrioventricular blocks and neurally mediated syncope). Proposals from AIAC (Italian Association of Arrhythmology and Cardiac Pacing). Issue 4 (April 2018)
- Main Title:
- Clinically oriented device programming in bradycardia patients
- Authors:
- Palmisano, Pietro
Ziacchi, Matteo
Biffi, Mauro
Ricci, Renato P.
Landolina, Maurizio
Zoni-Berisso, Massimo
Occhetta, Eraldo
Maglia, Giampiero
Botto, Gianluca
Padeletti, Luigi
Boriani, Giuseppe - Abstract:
- Abstract : N/A: The purpose of this two-part consensus document is to provide specific suggestions (based on an extensive literature review) on appropriate pacemaker setting in relation to patients' clinical features. In part 2, criteria for pacemaker choice and programming in atrioventricular blocks and neurally mediate syncope are proposed. The atrioventricular blocks can be paroxysmal or persistent, isolated or associated with sinus node disease. Neurally mediated syncope can be related to carotid sinus syndrome or cardioinhibitory vasovagal syncope. In sinus rhythm, with persistent atrioventricular block, we considered appropriate the activation of mode-switch algorithms, and algorithms for auto-adaptive management of the ventricular pacing output. If the atrioventricular block is paroxysmal, in addition to algorithms mentioned above, algorithms to maximize intrinsic atrioventricular conduction should be activated. When sinus node disease is associated with atrioventricular block, the activation of rate-responsive function in patients with chronotropic incompetence is appropriate. In permanent atrial fibrillation with atrioventricular block, algorithms for auto-adaptive management of the ventricular pacing output should be activated. If the atrioventricular block is persistent, the activation of rate-responsive function is appropriate. In carotid sinus syndrome, adequate rate hysteresis should be programmed. In vasovagal syncope, specialized sensing and pacing algorithmsAbstract : N/A: The purpose of this two-part consensus document is to provide specific suggestions (based on an extensive literature review) on appropriate pacemaker setting in relation to patients' clinical features. In part 2, criteria for pacemaker choice and programming in atrioventricular blocks and neurally mediate syncope are proposed. The atrioventricular blocks can be paroxysmal or persistent, isolated or associated with sinus node disease. Neurally mediated syncope can be related to carotid sinus syndrome or cardioinhibitory vasovagal syncope. In sinus rhythm, with persistent atrioventricular block, we considered appropriate the activation of mode-switch algorithms, and algorithms for auto-adaptive management of the ventricular pacing output. If the atrioventricular block is paroxysmal, in addition to algorithms mentioned above, algorithms to maximize intrinsic atrioventricular conduction should be activated. When sinus node disease is associated with atrioventricular block, the activation of rate-responsive function in patients with chronotropic incompetence is appropriate. In permanent atrial fibrillation with atrioventricular block, algorithms for auto-adaptive management of the ventricular pacing output should be activated. If the atrioventricular block is persistent, the activation of rate-responsive function is appropriate. In carotid sinus syndrome, adequate rate hysteresis should be programmed. In vasovagal syncope, specialized sensing and pacing algorithms designed for reflex syncope prevention should be activated. … (more)
- Is Part Of:
- Journal of cardiovascular medicine. Volume 19:Issue 4(2018:Apr.)
- Journal:
- Journal of cardiovascular medicine
- Issue:
- Volume 19:Issue 4(2018:Apr.)
- Issue Display:
- Volume 19, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 19
- Issue:
- 4
- Issue Sort Value:
- 2018-0019-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2018-04
- Subjects:
- atrial fibrillation -- atrioventricular block -- bifascicular block -- bradyarrhythmias -- carotid sinus syndrome -- device setting -- pacemaker -- pacing mode -- permanent pacing -- vasovagal syncope
Cardiology -- Periodicals
Cardiovascular system -- Diseases -- Periodicals
Cardiology -- Periodicals
Cardiovascular Diseases -- Periodicals
616.1005 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=01244665-000000000-00000 ↗
http://www.jcardiovascularmedicine.com/pt/re/jcm/home.htm ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.2459/JCM.0000000000000629 ↗
- Languages:
- English
- ISSNs:
- 1558-2027
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.867300
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6479.xml