Size effect on dose output in phantoms of x-ray tubes in medical x-ray imaging. (27th November 2017)
- Record Type:
- Journal Article
- Title:
- Size effect on dose output in phantoms of x-ray tubes in medical x-ray imaging. (27th November 2017)
- Main Title:
- Size effect on dose output in phantoms of x-ray tubes in medical x-ray imaging
- Authors:
- Zheng, Xiaoming
Nardi, Laura
Murray, Mark - Abstract:
- Abstract: The purpose of this work was to study how patient size affects the exponent of the power law relating dose to x-ray tube potential in clinical x-ray imaging using phantoms. Computed tomography (CT) dose phantoms of 16 cm and 32 cm in diameter were used to model children and adults respectively. A Fujifilm digital radiographic imaging system and a GE Discovery HD 750 CT scanner were employed to image both phantoms. For the Fujifilm radiographic imaging system, doses were measured at various locations of entrance, centre, exit as well as side locations of the phantoms using a PTW DIADOS diagnostic dosimeter. Standard procedure was employed for CT dose measurements for the GE CT scanner. X-ray exposures were varied by employing various tube voltages kVp and tube current-time product or mAs. This study found that the exponent in dose power law in phantoms increases as phantom size increases. Doses measured at body centres have lower uncertainties than those at x-ray exit surfaces. The power law exponent in phantoms was found to be a linear function of the depth from the entrance surface consistent with theoretical prediction and data from the literature. These results led to both kVp and body size dependent governing equations for the variations of imaging parameters in CT and radiographic imaging. These equations provide better estimates in guiding selection of optimal scan parameters in clinical x-ray imaging.
- Is Part Of:
- Biomedical physics & engineering express. Volume 3:Number 6(2017)
- Journal:
- Biomedical physics & engineering express
- Issue:
- Volume 3:Number 6(2017)
- Issue Display:
- Volume 3, Issue 6 (2017)
- Year:
- 2017
- Volume:
- 3
- Issue:
- 6
- Issue Sort Value:
- 2017-0003-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-11-27
- Subjects:
- dose output power -- size effect -- radiographic imaging -- computed tomography -- guiding equations
Medical physics -- Periodicals
Biophysics -- Periodicals
Biomedical engineering -- Periodicals
Medical sciences -- Periodicals
610.153 - Journal URLs:
- http://iopscience.iop.org/2057-1976/ ↗
http://www.iop.org/ ↗ - DOI:
- 10.1088/2057-1976/aa88f2 ↗
- Languages:
- English
- ISSNs:
- 2057-1976
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 6379.xml