Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T. Issue 3 (22nd September 2017)
- Record Type:
- Journal Article
- Title:
- Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T. Issue 3 (22nd September 2017)
- Main Title:
- Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T
- Authors:
- Wang, Jingkun
Fang, Yizhen
Li, Peihua
Lin, Huayue
Jin, Hongwei
Gao, Lei
Ge, Shengxiang
Zhang, Zhong‐Ying - Abstract:
- Abstract : Background: To evaluate the performance of a chemiluminescence detection kit for cardiac troponin T (cTnT). Methods: According to the "Guiding principles on performance analysis of diagnostic reagents in vitro" and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the detection limit, linear range, reportable range, accuracy, precision, cross reactivity, interference factors, and matrix effect of the kit, and compared these parameters with that of the commercial electrochemiluminescence detection kit for cTnT (Roche). Results: The minimum detection limit of the kit was 0.01 ng/mL. The linear range was 0.01 ng/mL‐25 ng/mL. The reportable range was from 0.01 ng/mL to 100 ng/mL. Precision within the batch was 2.9%‐6.4%, and precision between batches was 6.0%; the accuracy was good and the recovery rate reached 96.2%. The cross‐reaction test demonstrated that there was no reactivity between cTnT and a variety of troponins. Test results deviated by less than ±10% in the presence of hemoglobin ≤1000 μg/mL, bilirubin ≤250 μg/mL, triglycerides ≤11.25 mg/mL, and rheumatoid factor ≤206 U/mL, suggesting that kit results were not significantly interfered with these factors. Results from the matrix‐effect assessment revealed absence of a matrix effect in the tested serum samples. Correlation study revealed that the performance of the kit was very consistent with that of the Roche electrochemiluminescence detection kit (Kappa = 0.900, P < .001)Abstract : Background: To evaluate the performance of a chemiluminescence detection kit for cardiac troponin T (cTnT). Methods: According to the "Guiding principles on performance analysis of diagnostic reagents in vitro" and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the detection limit, linear range, reportable range, accuracy, precision, cross reactivity, interference factors, and matrix effect of the kit, and compared these parameters with that of the commercial electrochemiluminescence detection kit for cTnT (Roche). Results: The minimum detection limit of the kit was 0.01 ng/mL. The linear range was 0.01 ng/mL‐25 ng/mL. The reportable range was from 0.01 ng/mL to 100 ng/mL. Precision within the batch was 2.9%‐6.4%, and precision between batches was 6.0%; the accuracy was good and the recovery rate reached 96.2%. The cross‐reaction test demonstrated that there was no reactivity between cTnT and a variety of troponins. Test results deviated by less than ±10% in the presence of hemoglobin ≤1000 μg/mL, bilirubin ≤250 μg/mL, triglycerides ≤11.25 mg/mL, and rheumatoid factor ≤206 U/mL, suggesting that kit results were not significantly interfered with these factors. Results from the matrix‐effect assessment revealed absence of a matrix effect in the tested serum samples. Correlation study revealed that the performance of the kit was very consistent with that of the Roche electrochemiluminescence detection kit (Kappa = 0.900, P < .001) with a high correlation ( r = .903, P < .001) and a total matching rate of 95.00%. Conclusion: The performance of the evaluated chemiluminescence immunoquantitation kit for cTnT detection was acceptable in all tested parameters, fulfilling requirements for clinical applications. … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 32:Issue 3(2018)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 32:Issue 3(2018)
- Issue Display:
- Volume 32, Issue 3 (2018)
- Year:
- 2018
- Volume:
- 32
- Issue:
- 3
- Issue Sort Value:
- 2018-0032-0003-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2017-09-22
- Subjects:
- cardiac troponin T -- chemiluminescence immunoassay -- evaluation -- performance -- reagent
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.22311 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 6157.xml