Assessment of the variation in CT scanner performance (image quality and Hounsfield units) with scan parameters, for image optimisation in radiotherapy treatment planning. (January 2018)
- Record Type:
- Journal Article
- Title:
- Assessment of the variation in CT scanner performance (image quality and Hounsfield units) with scan parameters, for image optimisation in radiotherapy treatment planning. (January 2018)
- Main Title:
- Assessment of the variation in CT scanner performance (image quality and Hounsfield units) with scan parameters, for image optimisation in radiotherapy treatment planning
- Authors:
- Davis, Anne T.
Palmer, Antony L.
Pani, Silvia
Nisbet, Andrew - Abstract:
- Highlights: Test method to assess how CT parameters affect image quality and Hounsfield units. CNR and HCSR varied most with reconstruction algorithm, reconstruction FOV and mAs. Hounsfield units changed with acquisition FOV and reconstruction algorithm selected. The results will enable optimisation of CT protocols used for radiotherapy planning. Abstract: Purpose: To define a method and investigate how the adjustment of scan parameters affected the image quality and Hounsfield units (HUs) on a CT scanner used for radiotherapy treatment planning. A lack of similar investigations in the literature may be a contributing factor in the apparent reluctance to optimise radiotherapy CT protocols. Method: A Catphan phantom was used to assess how image quality on a Toshiba Aquilion LB scanner changed with scan parameters. Acquisition and reconstruction field-of-view (FOV), collimation, image slice thickness, effective mAs per rotation and reconstruction algorithm were varied. Changes were assessed for HUs of different materials, high contrast spatial resolution (HCSR), contrast-noise ratio (CNR), HU uniformity, scan direction low contrast and CT dose-index. Results: CNR and HCSR varied most with reconstruction algorithm, reconstruction FOV and effective mAs. Collimation, but not image slice width, had a significant effect on CT dose-index with narrower collimation giving higher doses. Dose increased with effective mAs. Highest HU differences were seen when changing reconstructionHighlights: Test method to assess how CT parameters affect image quality and Hounsfield units. CNR and HCSR varied most with reconstruction algorithm, reconstruction FOV and mAs. Hounsfield units changed with acquisition FOV and reconstruction algorithm selected. The results will enable optimisation of CT protocols used for radiotherapy planning. Abstract: Purpose: To define a method and investigate how the adjustment of scan parameters affected the image quality and Hounsfield units (HUs) on a CT scanner used for radiotherapy treatment planning. A lack of similar investigations in the literature may be a contributing factor in the apparent reluctance to optimise radiotherapy CT protocols. Method: A Catphan phantom was used to assess how image quality on a Toshiba Aquilion LB scanner changed with scan parameters. Acquisition and reconstruction field-of-view (FOV), collimation, image slice thickness, effective mAs per rotation and reconstruction algorithm were varied. Changes were assessed for HUs of different materials, high contrast spatial resolution (HCSR), contrast-noise ratio (CNR), HU uniformity, scan direction low contrast and CT dose-index. Results: CNR and HCSR varied most with reconstruction algorithm, reconstruction FOV and effective mAs. Collimation, but not image slice width, had a significant effect on CT dose-index with narrower collimation giving higher doses. Dose increased with effective mAs. Highest HU differences were seen when changing reconstruction algorithm: 56 HU for densities close to water and 117 HU for bone-like materials. Acquisition FOV affected the HUs but reconstruction FOV and effective mAs did not. Conclusions: All the scan parameters investigated affected the image quality metrics. Reconstruction algorithm, reconstruction FOV, collimation and effective mAs were most important. Reconstruction algorithm and acquisition FOV had significant effect on HU. The methodology is applicable to radiotherapy CT scanners when investigating image quality optimisation, prior to assessing the impact of scan protocol changes on clinical CT images and treatment plans. … (more)
- Is Part Of:
- Physica medica. Volume 45(2018)
- Journal:
- Physica medica
- Issue:
- Volume 45(2018)
- Issue Display:
- Volume 45, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 45
- Issue:
- 2018
- Issue Sort Value:
- 2018-0045-2018-0000
- Page Start:
- 59
- Page End:
- 64
- Publication Date:
- 2018-01
- Subjects:
- Radiotherapy -- CT -- Imaging -- Optimisation
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2017.11.036 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5857.xml