A novel high-resolution 2D silicon array detector for small field dosimetry with FFF photon beams. (January 2018)
- Record Type:
- Journal Article
- Title:
- A novel high-resolution 2D silicon array detector for small field dosimetry with FFF photon beams. (January 2018)
- Main Title:
- A novel high-resolution 2D silicon array detector for small field dosimetry with FFF photon beams
- Authors:
- Biasi, G.
Petasecca, M.
Guatelli, S.
Hardcastle, N.
Carolan, M.
Perevertaylo, V.
Kron, T.
Rosenfeld, A.B. - Abstract:
- Highlights: 2D monolithic silicon array detector with sub-millimetre spatial resolution. Characterization for small field relative dosimetry with FFF photon beams. OFs accurate within 3%. PDDs values accurate within 2%. In-plane, cross-plane, 2 diagonal dose profiles acquired simultaneously with OF. Abstract: Purpose: Flattening filter free (FFF) beams are increasingly being considered for stereotactic radiotherapy (SRT). For the first time, the performance of a monolithic silicon array detector under 6 and 10 MV FFF beams was evaluated. The dosimeter, named "Octa" and designed by the Centre for Medical Radiation Physics (CMRP), was tested also under flattened beams for comparison. Methods: Output factors (OFs), percentage depth-dose (PDD), dose profiles (DPs) and dose per pulse (DPP) dependence were investigated. Results were benchmarked against commercially available detectors for small field dosimetry. Results: The dosimeter was shown to be a 'correction-free' silicon array detector for OFs and PDD measurements for all the beam qualities investigated. Measured OFs were accurate within 3% and PDD values within 2% compared against the benchmarks. Cross-plane, in-plane and diagonal DPs were measured simultaneously with high spatial resolution (0.3 mm) and real time read-out. A DPP dependence (24% at 0.021 mGy/pulse relative to 0.278 mGy/pulse) was found and could be easily corrected for in the case of machine specific quality assurance applications. Conclusions: Results wereHighlights: 2D monolithic silicon array detector with sub-millimetre spatial resolution. Characterization for small field relative dosimetry with FFF photon beams. OFs accurate within 3%. PDDs values accurate within 2%. In-plane, cross-plane, 2 diagonal dose profiles acquired simultaneously with OF. Abstract: Purpose: Flattening filter free (FFF) beams are increasingly being considered for stereotactic radiotherapy (SRT). For the first time, the performance of a monolithic silicon array detector under 6 and 10 MV FFF beams was evaluated. The dosimeter, named "Octa" and designed by the Centre for Medical Radiation Physics (CMRP), was tested also under flattened beams for comparison. Methods: Output factors (OFs), percentage depth-dose (PDD), dose profiles (DPs) and dose per pulse (DPP) dependence were investigated. Results were benchmarked against commercially available detectors for small field dosimetry. Results: The dosimeter was shown to be a 'correction-free' silicon array detector for OFs and PDD measurements for all the beam qualities investigated. Measured OFs were accurate within 3% and PDD values within 2% compared against the benchmarks. Cross-plane, in-plane and diagonal DPs were measured simultaneously with high spatial resolution (0.3 mm) and real time read-out. A DPP dependence (24% at 0.021 mGy/pulse relative to 0.278 mGy/pulse) was found and could be easily corrected for in the case of machine specific quality assurance applications. Conclusions: Results were consistent with those for monolithic silicon array detectors designed by the CMRP and previously characterized under flattened beams only, supporting the robustness of this technology for relative dosimetry for a wide range of beam qualities and dose per pulses. In contrast to its predecessors, the design of the Octa offers an exhaustive high-resolution 2D dose map characterization, making it a unique real-time radiation detector for small field dosimetry for field sizes up to 3 cm side. … (more)
- Is Part Of:
- Physica medica. Volume 45(2018)
- Journal:
- Physica medica
- Issue:
- Volume 45(2018)
- Issue Display:
- Volume 45, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 45
- Issue:
- 2018
- Issue Sort Value:
- 2018-0045-2018-0000
- Page Start:
- 117
- Page End:
- 126
- Publication Date:
- 2018-01
- Subjects:
- SRT -- SBRT -- FFF -- Small field dosimetry QA -- 2D monolithic silicon array detector
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2017.12.010 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5856.xml