Accuracy of Noninvasive and Continuous Hemoglobin Measurement by Pulse Co-Oximetry During Preoperative Phlebotomy. (July 2014)
- Record Type:
- Journal Article
- Title:
- Accuracy of Noninvasive and Continuous Hemoglobin Measurement by Pulse Co-Oximetry During Preoperative Phlebotomy. (July 2014)
- Main Title:
- Accuracy of Noninvasive and Continuous Hemoglobin Measurement by Pulse Co-Oximetry During Preoperative Phlebotomy
- Authors:
- Dewhirst, Elisabeth
Naguib, Aymen
Winch, Peter
Rice, Julie
Galantowicz, Mark
McConnell, Patrick
Tobias, Joseph D. - Abstract:
- Background: In recent years, the continuous noninvasive hemoglobin measurement has been offered by devices using advanced pulse oximetry technology. Accuracy has been established in healthy adults as well as in surgical and intensive care unit patients but not in the setting of acute hemorrhage. In this study, we evaluated the accuracy of such a device in the clinical setting of preoperative phlebotomy thereby mimicking a scenario of acute blood loss. Methods: This prospective study included patients undergoing surgical repair of congenital heart disease (CHD) for whom preoperative phlebotomy was planned. Blood was removed after the induction of anesthesia and prior to the start of the surgical procedure. Replacement with crystalloid was guided by hemodynamic variables and cerebral oxygenation measured by near-infrared spectroscopy. Hemoglobin was measured by bedside whole blood analysis (total hemoglobin [tHb]) before and after phlebotomy, and concurrent measurements from the pulse co-oximeter (noninvasive, continuous, or spot-check testing of total hemoglobin [SpHb]) were recorded. Results: The study cohort included 45 patients ranging in age from 3 months to 50 years. Preoperative phlebotomy removed an average of 9.2 mL/kg of blood that was replaced with an average of 7.2 mL/kg of crystalloid. The pre- and postphlebotomy tHb values were 13.0 ± 1.9 and 12.4 ± 1.8 g/dL, respectively. The absolute difference between the tHb and SpHb (▵Hb) was 1.2 ± 0.1 g/dL. Bland-AltmanBackground: In recent years, the continuous noninvasive hemoglobin measurement has been offered by devices using advanced pulse oximetry technology. Accuracy has been established in healthy adults as well as in surgical and intensive care unit patients but not in the setting of acute hemorrhage. In this study, we evaluated the accuracy of such a device in the clinical setting of preoperative phlebotomy thereby mimicking a scenario of acute blood loss. Methods: This prospective study included patients undergoing surgical repair of congenital heart disease (CHD) for whom preoperative phlebotomy was planned. Blood was removed after the induction of anesthesia and prior to the start of the surgical procedure. Replacement with crystalloid was guided by hemodynamic variables and cerebral oxygenation measured by near-infrared spectroscopy. Hemoglobin was measured by bedside whole blood analysis (total hemoglobin [tHb]) before and after phlebotomy, and concurrent measurements from the pulse co-oximeter (noninvasive, continuous, or spot-check testing of total hemoglobin [SpHb]) were recorded. Results: The study cohort included 45 patients ranging in age from 3 months to 50 years. Preoperative phlebotomy removed an average of 9.2 mL/kg of blood that was replaced with an average of 7.2 mL/kg of crystalloid. The pre- and postphlebotomy tHb values were 13.0 ± 1.9 and 12.4 ± 1.8 g/dL, respectively. The absolute difference between the tHb and SpHb (▵Hb) was 1.2 ± 0.1 g/dL. Bland-Altman analysis revealed a bias of 0.1 g/dL, a precision of 1.5 g/dL, and 95% limits of agreement of −2.8 to 3.1 g/dL. In 52.2% of the sample sets, the SpHb was within 1 g/dL of the actual hemoglobin value (tHb), and in 80% of the sample sets, the SpHb was within 2 g/dL. No variation in the accuracy of the deviation was noted based on the patient's age, weight, or type of CHD (cyanotic versus acyanotic). Conclusion: The current study demonstrates that the accuracy of continuous, noninvasive hemoglobin measurement was not affected by acute blood loss simulated by preoperative phlebotomy. Although the device provided a clinically acceptable correlation with the actual hemoglobin value and offers the value of a continuous trend monitor, given the precision of the device, it does not appear that actual transfusion decisions can be based on the device alone. … (more)
- Is Part Of:
- Journal of intensive care medicine. Volume 29:Number 4(2014:Jul./Aug.)
- Journal:
- Journal of intensive care medicine
- Issue:
- Volume 29:Number 4(2014:Jul./Aug.)
- Issue Display:
- Volume 29, Issue 4 (2014)
- Year:
- 2014
- Volume:
- 29
- Issue:
- 4
- Issue Sort Value:
- 2014-0029-0004-0000
- Page Start:
- 238
- Page End:
- 242
- Publication Date:
- 2014-07
- Subjects:
- Non invasive hemoglobin monitoring -- Acute normovolemic hemodilution -- Monitoring and pediatric cardiac surgery
Critical care medicine -- Periodicals
Critical Care -- Periodicals
Soins intensifs -- Périodiques
Soins intensifs
Périodique électronique (Descripteur de forme)
Ressource Internet (Descripteur de forme)
616.02805 - Journal URLs:
- http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=0885-0666;screen=info;ECOIP ↗
http://jic.sagepub.com ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=jic ↗
http://www.sagepublications.com/ ↗ - DOI:
- 10.1177/0885066613485355 ↗
- Languages:
- English
- ISSNs:
- 0885-0666
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5846.xml