In vitro artifact assessment of an MR-compatible, microwave antenna device for percutaneous tumor ablation with fluoroscopic MRI-sequences. Issue 1 (2nd January 2018)
- Record Type:
- Journal Article
- Title:
- In vitro artifact assessment of an MR-compatible, microwave antenna device for percutaneous tumor ablation with fluoroscopic MRI-sequences. Issue 1 (2nd January 2018)
- Main Title:
- In vitro artifact assessment of an MR-compatible, microwave antenna device for percutaneous tumor ablation with fluoroscopic MRI-sequences
- Authors:
- Kessler, David-Emanuel
Weiss, Jakob
Rempp, Hansjörg
Pereira, Philippe L.
Nikolaou, Konstantin
Clasen, Stephan
Hoffmann, Rüdiger - Abstract:
- Abstract: Objective: To evaluate artifact configuration and diameters of a magnetic resonance (MR) compatible microwave (MW) applicator using near-realtime MR-fluoroscopic sequences for percutaneous tumor ablation procedures. Material and methods: Two MW applicators (14 G and 16 G) were tested in an ex-vivo phantom at 1.5 T with two 3 D fluoroscopic sequences: T1-weighted spoiled Gradient Echo (GRE) and T1/T2-weighted Steady State Free Precession (SSFP) sequence. Applicator orientation to main magnetic field (B0 ), slice orientation and phase encoding direction (PED) were systematically varied. The influence of these variables was assessed with ANOVA and post-hoc testing. Results: The artifact was homogenous along the whole length of both antennas with all tested parameters. The tip artifact diameter of the 16 G antenna measured 6.9 ± 1.0 mm, the shaft artifact diameter 8.6 ± 1.2 mm and the Tip Location Error (TLE) was 1.5 ± 1.2 mm.The tip artifact diameter of the 14 G antenna measured 7.7 ± 1.2 mm, the shaft artifact diameter 9.6 ± 1.5 mm and TLE was 1.6 ± 1.2 mm. Orientation to B0 had no statistically significant influence on tip artifact diameters (16 G: p = .55; 14 G: p = .07) or TLE (16 G: p = .93; 14 G: p = .26). GRE sequences slightly overestimated the antenna length with TLE(16 G) = 2.6 ± 0.5 mm and TLE(14 G) = 2.7 ± 0.7 mm. Conclusions: The MR-compatible MW applicator's artifact seems adequate with an acceptable TLE for safe applicator positioning duringAbstract: Objective: To evaluate artifact configuration and diameters of a magnetic resonance (MR) compatible microwave (MW) applicator using near-realtime MR-fluoroscopic sequences for percutaneous tumor ablation procedures. Material and methods: Two MW applicators (14 G and 16 G) were tested in an ex-vivo phantom at 1.5 T with two 3 D fluoroscopic sequences: T1-weighted spoiled Gradient Echo (GRE) and T1/T2-weighted Steady State Free Precession (SSFP) sequence. Applicator orientation to main magnetic field (B0 ), slice orientation and phase encoding direction (PED) were systematically varied. The influence of these variables was assessed with ANOVA and post-hoc testing. Results: The artifact was homogenous along the whole length of both antennas with all tested parameters. The tip artifact diameter of the 16 G antenna measured 6.9 ± 1.0 mm, the shaft artifact diameter 8.6 ± 1.2 mm and the Tip Location Error (TLE) was 1.5 ± 1.2 mm.The tip artifact diameter of the 14 G antenna measured 7.7 ± 1.2 mm, the shaft artifact diameter 9.6 ± 1.5 mm and TLE was 1.6 ± 1.2 mm. Orientation to B0 had no statistically significant influence on tip artifact diameters (16 G: p = .55; 14 G: p = .07) or TLE (16 G: p = .93; 14 G: p = .26). GRE sequences slightly overestimated the antenna length with TLE(16 G) = 2.6 ± 0.5 mm and TLE(14 G) = 2.7 ± 0.7 mm. Conclusions: The MR-compatible MW applicator's artifact seems adequate with an acceptable TLE for safe applicator positioning during near-realtime fluoroscopic MR-guidance. … (more)
- Is Part Of:
- Minimally invasive therapy & allied technologies. Volume 27:Issue 1(2018)
- Journal:
- Minimally invasive therapy & allied technologies
- Issue:
- Volume 27:Issue 1(2018)
- Issue Display:
- Volume 27, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 27
- Issue:
- 1
- Issue Sort Value:
- 2018-0027-0001-0000
- Page Start:
- 60
- Page End:
- 68
- Publication Date:
- 2018-01-02
- Subjects:
- Microwave ablation -- MR-guidance -- MR-fluoroscopy -- thermal ablation
Endoscopy -- Periodicals
Interventional radiology -- Periodicals
Endoscopic surgery -- Periodicals
617.05 - Journal URLs:
- http://informahealthcare.com/loi/mit ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/13645706.2017.1414062 ↗
- Languages:
- English
- ISSNs:
- 1364-5706
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5797.714000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5729.xml