Evaluation of clinically applied treatment beams with respect to bunker shielding parameters for a Cyberknife M6. (10th November 2017)
- Record Type:
- Journal Article
- Title:
- Evaluation of clinically applied treatment beams with respect to bunker shielding parameters for a Cyberknife M6. (10th November 2017)
- Main Title:
- Evaluation of clinically applied treatment beams with respect to bunker shielding parameters for a Cyberknife M6
- Authors:
- Henzen, Dominik
Schmidhalter, Daniel
Zanella, Claudia Christina
Volken, Werner
Mackeprang, Paul‐Henry
Malthaner, Marco
Fix, Michael Karl
Manser, Peter - Abstract:
- Abstract: Compared to a conventional linear accelerator, the Cyberknife (CK) is a unique system with respect to radiation protection shielding and the variety and number of non‐coplanar beams are two key components regarding this aspect. In this work, a framework to assess the direction distribution and modulation factor (MF) of clinically applied treatment beams of a CyberKnife M6 is developed. Database filtering options allow studying the influence of different parameters such as collimator types, treatment sites or different bunker sizes. A distribution of monitor units (MU) is generated by projecting treatment beams onto the walls, floor and ceiling of the CyberKnife bunker. This distribution is found to be highly heterogeneous and depending, among other parameters, on the bunker size. For our bunker design, 10%–13% of the MUs are delivered to the right and left wall, each. The floor receives more than 64% of the applied MUs, while the wall behind the patient's head is not hit by primary treatment beams. Between 0% and 5% of the total MUs are delivered to the wall at the patient's feet. This number highly depends on the treatment site, e.g., for extracranial patients no beams hit that wall. Collimator choice was found to have minor influence on the distribution of MUs. On the other hand, the MF depends on the collimator type as well as on the treatment site. The MFs (delivered MU/prescribed dose) for all treatments, all MLC treatments, cranial and extracranial treatmentsAbstract: Compared to a conventional linear accelerator, the Cyberknife (CK) is a unique system with respect to radiation protection shielding and the variety and number of non‐coplanar beams are two key components regarding this aspect. In this work, a framework to assess the direction distribution and modulation factor (MF) of clinically applied treatment beams of a CyberKnife M6 is developed. Database filtering options allow studying the influence of different parameters such as collimator types, treatment sites or different bunker sizes. A distribution of monitor units (MU) is generated by projecting treatment beams onto the walls, floor and ceiling of the CyberKnife bunker. This distribution is found to be highly heterogeneous and depending, among other parameters, on the bunker size. For our bunker design, 10%–13% of the MUs are delivered to the right and left wall, each. The floor receives more than 64% of the applied MUs, while the wall behind the patient's head is not hit by primary treatment beams. Between 0% and 5% of the total MUs are delivered to the wall at the patient's feet. This number highly depends on the treatment site, e.g., for extracranial patients no beams hit that wall. Collimator choice was found to have minor influence on the distribution of MUs. On the other hand, the MF depends on the collimator type as well as on the treatment site. The MFs (delivered MU/prescribed dose) for all treatments, all MLC treatments, cranial and extracranial treatments are 8.3, 6.4, 7.7, and 9.9 MU/cGy, respectively. The developed framework allows assessing and monitoring important parameters regarding radiation protection of a CK‐M6 using the actually applied treatment beams. Furthermore, it enables evaluating different clinical and constructional situations using the filtering options. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 19:Number 1(2018)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 19:Number 1(2018)
- Issue Display:
- Volume 19, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2018-0019-0001-0000
- Page Start:
- 243
- Page End:
- 249
- Publication Date:
- 2017-11-10
- Subjects:
- Cyberknife -- radiation protection -- SBRT
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12215 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5637.xml