Hemodynamic Evaluation of a Biological and Mechanical Aortic Valve Prosthesis Using Patient‐Specific MRI‐Based CFD. Issue 1 (29th August 2017)
- Record Type:
- Journal Article
- Title:
- Hemodynamic Evaluation of a Biological and Mechanical Aortic Valve Prosthesis Using Patient‐Specific MRI‐Based CFD. Issue 1 (29th August 2017)
- Main Title:
- Hemodynamic Evaluation of a Biological and Mechanical Aortic Valve Prosthesis Using Patient‐Specific MRI‐Based CFD
- Authors:
- Hellmeier, Florian
Nordmeyer, Sarah
Yevtushenko, Pavlo
Bruening, Jan
Berger, Felix
Kuehne, Titus
Goubergrits, Leonid
Kelm, Marcus - Abstract:
- Abstract: Modeling different treatment options before a procedure is performed is a promising approach for surgical decision making and patient care in heart valve disease. This study investigated the hemodynamic impact of different prostheses through patient‐specific MRI‐based CFD simulations. Ten time‐resolved MRI data sets with and without velocity encoding were obtained to reconstruct the aorta and set hemodynamic boundary conditions for simulations. Aortic hemodynamics after virtual valve replacement with a biological and mechanical valve prosthesis were investigated. Wall shear stress (WSS), secondary flow degree (SFD), transvalvular pressure drop (TPD), turbulent kinetic energy (TKE), and normalized flow displacement (NFD) were evaluated to characterize valve‐induced hemodynamics. The biological prostheses induced significantly higher WSS (medians: 9.3 vs. 8.6 Pa, P = 0.027) and SFD (means: 0.78 vs. 0.49, P = 0.002) in the ascending aorta, TPD (medians: 11.4 vs. 2.7 mm Hg, P = 0.002), TKE (means: 400 vs. 283 cm 2 /s 2, P = 0.037), and NFD (means: 0.0994 vs. 0.0607, P = 0.020) than the mechanical prostheses. The differences between the prosthesis types showed great inter‐patient variability, however. Given this variability, a patient‐specific evaluation is warranted. In conclusion, MRI‐based CFD offers an opportunity to assess the interactions between prosthesis and patient‐specific boundary conditions, which may help in optimizing surgical decision making andAbstract: Modeling different treatment options before a procedure is performed is a promising approach for surgical decision making and patient care in heart valve disease. This study investigated the hemodynamic impact of different prostheses through patient‐specific MRI‐based CFD simulations. Ten time‐resolved MRI data sets with and without velocity encoding were obtained to reconstruct the aorta and set hemodynamic boundary conditions for simulations. Aortic hemodynamics after virtual valve replacement with a biological and mechanical valve prosthesis were investigated. Wall shear stress (WSS), secondary flow degree (SFD), transvalvular pressure drop (TPD), turbulent kinetic energy (TKE), and normalized flow displacement (NFD) were evaluated to characterize valve‐induced hemodynamics. The biological prostheses induced significantly higher WSS (medians: 9.3 vs. 8.6 Pa, P = 0.027) and SFD (means: 0.78 vs. 0.49, P = 0.002) in the ascending aorta, TPD (medians: 11.4 vs. 2.7 mm Hg, P = 0.002), TKE (means: 400 vs. 283 cm 2 /s 2, P = 0.037), and NFD (means: 0.0994 vs. 0.0607, P = 0.020) than the mechanical prostheses. The differences between the prosthesis types showed great inter‐patient variability, however. Given this variability, a patient‐specific evaluation is warranted. In conclusion, MRI‐based CFD offers an opportunity to assess the interactions between prosthesis and patient‐specific boundary conditions, which may help in optimizing surgical decision making and providing additional guidance to clinicians. … (more)
- Is Part Of:
- Artificial organs. Volume 42:Issue 1(2018)
- Journal:
- Artificial organs
- Issue:
- Volume 42:Issue 1(2018)
- Issue Display:
- Volume 42, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 42
- Issue:
- 1
- Issue Sort Value:
- 2018-0042-0001-0000
- Page Start:
- 49
- Page End:
- 57
- Publication Date:
- 2017-08-29
- Subjects:
- Patient‐specific simulation -- —Aortic hemodynamics -- —Aortic valve replacement -- —Virtual intervention -- —4D flow MRI
Artificial organs -- Periodicals
617.956 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1525-1594 ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=aor ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1111/aor.12955 ↗
- Languages:
- English
- ISSNs:
- 0160-564X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1735.052000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5599.xml