Circuit Oxygenator Contributes to Extracorporeal Membrane Oxygenation–Induced Hemolysis. Issue 2 (March 2015)
- Record Type:
- Journal Article
- Title:
- Circuit Oxygenator Contributes to Extracorporeal Membrane Oxygenation–Induced Hemolysis. Issue 2 (March 2015)
- Main Title:
- Circuit Oxygenator Contributes to Extracorporeal Membrane Oxygenation–Induced Hemolysis
- Authors:
- Williams, Duane C.
Turi, Jennifer L.
Hornik, Christoph P.
Bonadonna, Desiree K.
Williford, Walter L.
Walczak, Richard J.
Watt, Kevin M.
Cheifetz, Ira M. - Abstract:
- Abstract : Hemolysis can occur as a consequence of extracorporeal membrane oxygenation (ECMO) and is associated with increased mortality and morbidity. Shear stress generated by flow through the circuit and oxygenator is believed to cause ECMO-induced hemolysis. We hypothesize that either a smaller dimension oxygenator or an in-line hemofilter will increase ECMO-associated hemolysis. Circuits were configured with a Quadrox-D Adult oxygenator (surface area 1.8 m 2 ), Quadrox-iD Pediatric oxygenator (surface area 0.8 m 2 ), or Quadrox-D Adult oxygenator with an in-line hemofilter (N = 4) and ran for 6 hours. Samples were collected hourly from the ECMO circuit and a time-based hemolysis control. Plasma hemoglobin levels were assayed. Circuit-induced hemolysis at each time point was defined as the change in plasma hemoglobin standardized to the time-based hemolysis control. Plasma hemoglobin increased with the use of the smaller dimension pediatric oxygenator as compared with the adult oxygenator when controlling for ECMO run time ( p = 0.02). Furthermore, there was a greater pressure gradient with the smaller dimension pediatric oxygenator ( p < 0.05). Plasma hemoglobin did not change with the addition of the in-line hemofilter. The use of a smaller dimension pediatric oxygenator resulted in greater hemolysis and a higher pressure gradient. This may indicate that the increased shear forces augment ECMO-induced hemolysis.
- Is Part Of:
- ASAIO journal. Volume 61:Issue 2(2015)
- Journal:
- ASAIO journal
- Issue:
- Volume 61:Issue 2(2015)
- Issue Display:
- Volume 61, Issue 2 (2015)
- Year:
- 2015
- Volume:
- 61
- Issue:
- 2
- Issue Sort Value:
- 2015-0061-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-03
- Subjects:
- pediatric -- oxygenator -- extracorporeal membrane oxygenation -- hemolysis
Artificial organs -- Periodicals
617 - Journal URLs:
- http://journals.lww.com/asaiojournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAT.0000000000000173 ↗
- Languages:
- English
- ISSNs:
- 1058-2916
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1738.840500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5498.xml