Performance of Liaison XL automated immunoassay platform for blood‐borne infection screening on hepatitis B, hepatitis C, HIV 1/2, HTLV 1/2 and Treponema pallidum serological markers. Issue 2 (17th March 2015)
- Record Type:
- Journal Article
- Title:
- Performance of Liaison XL automated immunoassay platform for blood‐borne infection screening on hepatitis B, hepatitis C, HIV 1/2, HTLV 1/2 and Treponema pallidum serological markers. Issue 2 (17th March 2015)
- Main Title:
- Performance of Liaison XL automated immunoassay platform for blood‐borne infection screening on hepatitis B, hepatitis C, HIV 1/2, HTLV 1/2 and Treponema pallidum serological markers
- Authors:
- Malm, K.
Kragsbjerg, E.
Andersson, S. - Abstract:
- SUMMARY: Objectives: Aim of the study was to evaluate performance of a new fully automated platform, DiaSorin‐LIAISON® XL (DiaSorin S.p.A, Vercelli, Italy), in blood donor screening, specifically for hepatitis B surface antigen (HBsAg), hepatitis B core antibodies (anti‐HBc), hepatitis C antibodies (anti‐HCV), HIV p24 antigen, HIV antibodies, human T‐lymphotropic virus types 1 and 2 (HTLV‐1/2) and Treponema pallidum antibodies. Background: In screening for such viral and bacteriological blood‐borne infections, sensitivity and specificity are of utmost importance. Methods: Sensitivity was evaluated using selected panels of samples previously analysed on the Abbott Architect immunoanalyser (Abbott Laboratories, Abbott Park, IL, USA) – the gold standard for this evaluation. These samples were confirmed positive for HBsAg, anti‐HBc, anti‐HCV, HIV Ag/Ab, anti‐HTLV‐1/2 and antibodies to T. pallidum, respectively. Specificity analysis was assessed by analysing blood donor samples previously run on the Architect platform and found non‐reactive for each marker. A total of 1·100 donor samples (both new and regular donors) were tested. Previously, non‐specific reactive samples were also run for every tested marker, as well as samples with autoimmune antibodies and antibodies to other infections. Results: Three hundred seventy‐eight samples positive for the tested markers (HBsAg n = 51, anti‐HBc n = 52, anti‐HCV n = 75, anti‐Treponema n = 55, anti‐HIV‐1 n = 79, anti‐HIV‐2 n = 25,SUMMARY: Objectives: Aim of the study was to evaluate performance of a new fully automated platform, DiaSorin‐LIAISON® XL (DiaSorin S.p.A, Vercelli, Italy), in blood donor screening, specifically for hepatitis B surface antigen (HBsAg), hepatitis B core antibodies (anti‐HBc), hepatitis C antibodies (anti‐HCV), HIV p24 antigen, HIV antibodies, human T‐lymphotropic virus types 1 and 2 (HTLV‐1/2) and Treponema pallidum antibodies. Background: In screening for such viral and bacteriological blood‐borne infections, sensitivity and specificity are of utmost importance. Methods: Sensitivity was evaluated using selected panels of samples previously analysed on the Abbott Architect immunoanalyser (Abbott Laboratories, Abbott Park, IL, USA) – the gold standard for this evaluation. These samples were confirmed positive for HBsAg, anti‐HBc, anti‐HCV, HIV Ag/Ab, anti‐HTLV‐1/2 and antibodies to T. pallidum, respectively. Specificity analysis was assessed by analysing blood donor samples previously run on the Architect platform and found non‐reactive for each marker. A total of 1·100 donor samples (both new and regular donors) were tested. Previously, non‐specific reactive samples were also run for every tested marker, as well as samples with autoimmune antibodies and antibodies to other infections. Results: Three hundred seventy‐eight samples positive for the tested markers (HBsAg n = 51, anti‐HBc n = 52, anti‐HCV n = 75, anti‐Treponema n = 55, anti‐HIV‐1 n = 79, anti‐HIV‐2 n = 25, anti‐HIV 1/2 n = 3, anti‐HTLV‐1 n = 28 anti‐HTLV‐2 n = 10) were tested and found positive, suggesting a high sensitivity. A number of 342–1100 negative blood donors (depending on marker) have been tested, with very good specificity for the markers tested, ranging between 99·5 and 100%, respectively. Conclusions: The LIAISON® XL platform demonstrated very high sensitivity for the markers tested and the specificity necessary to fulfil the stringent requirements for blood donor screening. … (more)
- Is Part Of:
- Transfusion medicine. Volume 25:Issue 2(2015)
- Journal:
- Transfusion medicine
- Issue:
- Volume 25:Issue 2(2015)
- Issue Display:
- Volume 25, Issue 2 (2015)
- Year:
- 2015
- Volume:
- 25
- Issue:
- 2
- Issue Sort Value:
- 2015-0025-0002-0000
- Page Start:
- 101
- Page End:
- 105
- Publication Date:
- 2015-03-17
- Subjects:
- blood‐borne infections -- immunoassay -- Liaison XL -- sensitivity -- specificity
Blood -- Transfusion -- Periodicals
615.39 - Journal URLs:
- http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=tme ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-3148 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/tme.12181 ↗
- Languages:
- English
- ISSNs:
- 0958-7578
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9020.706000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5450.xml