Development and Evaluation of Heartbeat: A Machine Perfusion Heart Preservation System. Issue 11 (11th August 2017)
- Record Type:
- Journal Article
- Title:
- Development and Evaluation of Heartbeat: A Machine Perfusion Heart Preservation System. Issue 11 (11th August 2017)
- Main Title:
- Development and Evaluation of Heartbeat: A Machine Perfusion Heart Preservation System
- Authors:
- Li, Yongnan
Zeng, Qingdong
Liu, Gang
Du, Junzhe
Gao, Bingren
Wang, Wei
Zheng, Zhe
Hu, Shengshou
Ji, Bingyang - Other Names:
- Azevedo Coste Christine guestEditor.
Popovic Milos guestEditor.
Mayr Winfried guestEditor. - Abstract:
- Abstract: Static cold storage is accompanied with a partial safe ischemic interval for donor hearts. In this current study, a machine perfusion system was built to provide a better preservation for the donor heart and assessment for myocardial function. Chinese mini‐swine (weight 30–35 kg, n = 16) were randomly divided into HTK, Celsior, and Heartbeat groups. All donor hearts were respectively preserved for 8 hours under static cold storage or machine perfusion. The perfusion solution is aimed to maintain its homeostasis based on monitoring the Heartbeat group. The ultrastructure of myocardium suggests better myocardial protection in the Heartbeat group compared with HTK or Celsior‐preserved hearts. The myocardial and coronary artery structural and functional integrity was evaluated by immunofluorescence and Western blots in the Heartbeat. In the Heartbeat group, donor hearts maintained a high adenosine triphosphate level. Bcl‐2 and Beclin‐1 protein demonstrates high expression in the Celsior group. The Heartbeat system can be used to preserve donor hearts, and it could guarantee the myocardial and endothelial function of hearts during machine perfusion. Translating Heartbeat into clinical practice, it is such as to impact on donor heart preservation for cardiac transplantation.
- Is Part Of:
- Artificial organs. Volume 41:Issue 11(2017)
- Journal:
- Artificial organs
- Issue:
- Volume 41:Issue 11(2017)
- Issue Display:
- Volume 41, Issue 11 (2017)
- Year:
- 2017
- Volume:
- 41
- Issue:
- 11
- Issue Sort Value:
- 2017-0041-0011-0000
- Page Start:
- E240
- Page End:
- E250
- Publication Date:
- 2017-08-11
- Subjects:
- Heart transplantation -- —Organ preservation -- —Machine perfusion
Artificial organs -- Periodicals
617.956 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1525-1594 ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=aor ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1111/aor.12867 ↗
- Languages:
- English
- ISSNs:
- 0160-564X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1735.052000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5439.xml