Subretinal Visual Implant Alpha IMS – Clinical trial interim report. (June 2015)
- Record Type:
- Journal Article
- Title:
- Subretinal Visual Implant Alpha IMS – Clinical trial interim report. (June 2015)
- Main Title:
- Subretinal Visual Implant Alpha IMS – Clinical trial interim report
- Authors:
- Stingl, Katarina
Bartz-Schmidt, Karl Ulrich
Besch, Dorothea
Chee, Caroline K.
Cottriall, Charles L.
Gekeler, Florian
Groppe, Markus
Jackson, Timothy L.
MacLaren, Robert E.
Koitschev, Assen
Kusnyerik, Akos
Neffendorf, James
Nemeth, Janos
Naeem, Mohamed Adheem Naser
Peters, Tobias
Ramsden, James D.
Sachs, Helmut
Simpson, Andrew
Singh, Mandeep S.
Wilhelm, Barbara
Wong, David
Zrenner, Eberhart - Abstract:
- Highlights: Retina Implant Alpha IMS was applied in a clinical trial in 29 blind patients. Visual acuity measured with Landolt C-rings reached up to 20/546. Grating acuity reached up to 3.3 cycles per degree. Four patients could read letters with the implant. 13 participants (45%) reported useful visual functions in their daily life. Abstract: A subretinal visual implant (Alpha IMS, Retina Implant AG, Reutlingen, Germany) was implanted in 29 blind participants with outer retinal degeneration in an international multicenter clinical trial. Primary efficacy endpoints of the study protocol were a significant improvement of activities of daily living and mobility to be assessed by activities of daily living tasks, recognition tasks, mobility, or a combination thereof. Secondary efficacy endpoints were a significant improvement of visual acuity/light perception and/or object recognition (clinicaltrials.gov, NCT01024803). During up to 12 months observation time twenty-one participants (72%) reached the primary endpoints, of which thirteen participants (45%) reported restoration of visual function which they use in daily life. Additionally, detection, localization, and identification of objects were significantly better with the implant power switched on in the first 3 months. Twenty-five participants (86%) reached the secondary endpoints. Measurable grating acuity was up to 3.3 cycles per degree, visual acuities using standardized Landolt C-rings were 20/2000, 20/2000, 20/606 andHighlights: Retina Implant Alpha IMS was applied in a clinical trial in 29 blind patients. Visual acuity measured with Landolt C-rings reached up to 20/546. Grating acuity reached up to 3.3 cycles per degree. Four patients could read letters with the implant. 13 participants (45%) reported useful visual functions in their daily life. Abstract: A subretinal visual implant (Alpha IMS, Retina Implant AG, Reutlingen, Germany) was implanted in 29 blind participants with outer retinal degeneration in an international multicenter clinical trial. Primary efficacy endpoints of the study protocol were a significant improvement of activities of daily living and mobility to be assessed by activities of daily living tasks, recognition tasks, mobility, or a combination thereof. Secondary efficacy endpoints were a significant improvement of visual acuity/light perception and/or object recognition (clinicaltrials.gov, NCT01024803). During up to 12 months observation time twenty-one participants (72%) reached the primary endpoints, of which thirteen participants (45%) reported restoration of visual function which they use in daily life. Additionally, detection, localization, and identification of objects were significantly better with the implant power switched on in the first 3 months. Twenty-five participants (86%) reached the secondary endpoints. Measurable grating acuity was up to 3.3 cycles per degree, visual acuities using standardized Landolt C-rings were 20/2000, 20/2000, 20/606 and 20/546. Maximal correct motion perception ranged from 3 to 35 degrees per second. These results show that subretinal implants can restore very-low-vision or low vision in blind (light perception or less) patients with end-stage hereditary retinal degenerations. … (more)
- Is Part Of:
- Vision research. Volume 111(2015)Part B
- Journal:
- Vision research
- Issue:
- Volume 111(2015)Part B
- Issue Display:
- Volume 111, Issue 2015 (2015)
- Year:
- 2015
- Volume:
- 111
- Issue:
- 2015
- Issue Sort Value:
- 2015-0111-2015-0000
- Page Start:
- 149
- Page End:
- 160
- Publication Date:
- 2015-06
- Subjects:
- Subretinal Implant Alpha IMS -- Neuroprosthetics -- Retinitis pigmentosa -- Artificial vision -- Hereditary retinal diseases -- Photoreceptor degeneration
Vision -- Periodicals
573.88 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00426989 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.visres.2015.03.001 ↗
- Languages:
- English
- ISSNs:
- 0042-6989
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9240.925000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5425.xml