Editor's Choice – Occurrence and Classification of Proximal Type I Endoleaks After EndoVascular Aneurysm Sealing Using the Nellix™ Device. (December 2017)
- Record Type:
- Journal Article
- Title:
- Editor's Choice – Occurrence and Classification of Proximal Type I Endoleaks After EndoVascular Aneurysm Sealing Using the Nellix™ Device. (December 2017)
- Main Title:
- Editor's Choice – Occurrence and Classification of Proximal Type I Endoleaks After EndoVascular Aneurysm Sealing Using the Nellix™ Device
- Authors:
- van den Ham, Leo H.
Holden, Andrew
Savlovskis, Janis
Witterbottom, Andrew
Ouriel, Kenneth
Reijnen, Michel M.P.J. - Abstract:
- Abstract : Objective/Background: Proximal type I endoleaks are associated with abdominal aortic aneurysm (AAA) growth and rupture and necessitate repair. The Nellix™ EndoVascular Aneurysm Sealing (EVAS) system is a unique approach to AAA repair, where the appearance and treatment of endoleaks is also different. This study aimed to analyse and categorise proximal endoleaks in an EVAS treated cohort. Methods: All patients, treated from February 2013 to December 2015, in 15 experienced EVAS centres, presenting with proximal endoleak were included. Computed tomography scans were analysed by a core laboratory. A consensus meeting was organised to discuss and qualify each case for selection, technical aspects, and possible causes of the endoleak. Endoleaks were classified using a novel classification system for EVAS. Results: During the study period 1851 patients were treated using EVAS at 15 centres and followed for a median of 494 ± 283 days. Among these, 58 cases (3.1%) developed a proximal endoleak (1.5% early and 1.7% late); of these, 84% of 58 patients were treated outside the original and 96% outside the current, refined, instructions for use. Low stent positioning was the most likely cause in 44.6%, a hostile anatomy in 16.1%, and a combination of both in 33.9%. Treatment, by embolisation or proximal extension, was performed in 47% of cases, with a technical success of 97%. Conclusion: The overall incidence of proximal endoleak after EVAS is 3.1% after a mean follow-upAbstract : Objective/Background: Proximal type I endoleaks are associated with abdominal aortic aneurysm (AAA) growth and rupture and necessitate repair. The Nellix™ EndoVascular Aneurysm Sealing (EVAS) system is a unique approach to AAA repair, where the appearance and treatment of endoleaks is also different. This study aimed to analyse and categorise proximal endoleaks in an EVAS treated cohort. Methods: All patients, treated from February 2013 to December 2015, in 15 experienced EVAS centres, presenting with proximal endoleak were included. Computed tomography scans were analysed by a core laboratory. A consensus meeting was organised to discuss and qualify each case for selection, technical aspects, and possible causes of the endoleak. Endoleaks were classified using a novel classification system for EVAS. Results: During the study period 1851 patients were treated using EVAS at 15 centres and followed for a median of 494 ± 283 days. Among these, 58 cases (3.1%) developed a proximal endoleak (1.5% early and 1.7% late); of these, 84% of 58 patients were treated outside the original and 96% outside the current, refined, instructions for use. Low stent positioning was the most likely cause in 44.6%, a hostile anatomy in 16.1%, and a combination of both in 33.9%. Treatment, by embolisation or proximal extension, was performed in 47% of cases, with a technical success of 97%. Conclusion: The overall incidence of proximal endoleak after EVAS is 3.1% after a mean follow-up period of 16 months, with 1.5% occurring within 30 days. Their occurrence is related to patient selection and stent positioning. Early detection and classification is crucial to avoid the potential of sac rupture. … (more)
- Is Part Of:
- European journal of vascular and endovascular surgery. Volume 54:Number 6(2017)
- Journal:
- European journal of vascular and endovascular surgery
- Issue:
- Volume 54:Number 6(2017)
- Issue Display:
- Volume 54, Issue 6 (2017)
- Year:
- 2017
- Volume:
- 54
- Issue:
- 6
- Issue Sort Value:
- 2017-0054-0006-0000
- Page Start:
- 729
- Page End:
- 736
- Publication Date:
- 2017-12
- Subjects:
- Classification -- EndoVascular aneurysm sealing -- Endoleak -- EVAS -- Incidence -- Nellix
Blood-vessels -- Endoscopic surgery -- Periodicals
Blood-vessels -- Surgery -- Periodicals
Vascular Surgical Procedures -- Periodicals
Vascular Surgical Procedures -- methods -- Periodicals
Vaisseaux sanguins -- Chirurgie -- Périodiques
Vaisseaux sanguins -- Chirurgie endoscopique -- Périodiques
Blood-vessels -- Endoscopic surgery
Blood-vessels -- Surgery
Endoscopy
Electronic journals
Periodicals
Electronic journals
617.413005 - Journal URLs:
- http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=1078-5884;screen=info;ECOIP ↗
http://www.harcourt-international.com/journals/ejvs/ ↗
http://www.harcourt-international.com/journals/ejvx/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/10785884 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/10785884 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ejvs.2017.09.016 ↗
- Languages:
- English
- ISSNs:
- 1078-5884
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.747280
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5397.xml