Investigating steam penetration using thermometric methods in dental handpieces with narrow internal lumens during sterilizing processes with non-vacuum or vacuum processes. Issue 4 (December 2017)
- Record Type:
- Journal Article
- Title:
- Investigating steam penetration using thermometric methods in dental handpieces with narrow internal lumens during sterilizing processes with non-vacuum or vacuum processes. Issue 4 (December 2017)
- Main Title:
- Investigating steam penetration using thermometric methods in dental handpieces with narrow internal lumens during sterilizing processes with non-vacuum or vacuum processes
- Authors:
- Winter, S.
Smith, A.
Lappin, D.
McDonagh, G.
Kirk, B. - Abstract:
- Summary: Background: Dental handpieces are required to be sterilized between patient use. Vacuum steam sterilization processes with fractionated pre/post-vacuum phases or unique cycles for specified medical devices are required for hollow instruments with internal lumens to assure successful air removal. Entrapped air will compromise achievement of required sterilization conditions. Many countries and professional organizations still advocate non-vacuum sterilization processes for these devices. Aim: To investigate non-vacuum downward/gravity displacement, type-N steam sterilization of dental handpieces, using thermometric methods to measure time to achieve sterilization temperature at different handpiece locations. Methods: Measurements at different positions within air turbines were undertaken with thermocouples and data loggers. Two examples of widely used UK benchtop steam sterilizers were tested: a non-vacuum benchtop sterilizer (Little Sister 3; Eschmann, Lancing, UK) and a vacuum benchtop sterilizer (Lisa; W&H, Bürmoos, Austria). Each sterilizer cycle was completed with three handpieces and each cycle in triplicate. Findings: A total of 140 measurements inside dental handpiece lumens were recorded. The non-vacuum process failed (time range: 0–150 s) to reliably achieve sterilization temperatures within the time limit specified by the international standard (15 s equilibration time). The measurement point at the base of the handpiece failed in all test runs ( N = 9)Summary: Background: Dental handpieces are required to be sterilized between patient use. Vacuum steam sterilization processes with fractionated pre/post-vacuum phases or unique cycles for specified medical devices are required for hollow instruments with internal lumens to assure successful air removal. Entrapped air will compromise achievement of required sterilization conditions. Many countries and professional organizations still advocate non-vacuum sterilization processes for these devices. Aim: To investigate non-vacuum downward/gravity displacement, type-N steam sterilization of dental handpieces, using thermometric methods to measure time to achieve sterilization temperature at different handpiece locations. Methods: Measurements at different positions within air turbines were undertaken with thermocouples and data loggers. Two examples of widely used UK benchtop steam sterilizers were tested: a non-vacuum benchtop sterilizer (Little Sister 3; Eschmann, Lancing, UK) and a vacuum benchtop sterilizer (Lisa; W&H, Bürmoos, Austria). Each sterilizer cycle was completed with three handpieces and each cycle in triplicate. Findings: A total of 140 measurements inside dental handpiece lumens were recorded. The non-vacuum process failed (time range: 0–150 s) to reliably achieve sterilization temperatures within the time limit specified by the international standard (15 s equilibration time). The measurement point at the base of the handpiece failed in all test runs ( N = 9) to meet the standard. No failures were detected with the vacuum steam sterilization type B process with fractionated pre-vacuum and post-vacuum phases. Conclusion: Non-vacuum downward/gravity displacement, type-N steam sterilization processes are unreliable in achieving sterilization conditions inside dental handpieces, and the base of the handpiece is the site most likely to fail. … (more)
- Is Part Of:
- Journal of hospital infection. Volume 97:Issue 4(2017)
- Journal:
- Journal of hospital infection
- Issue:
- Volume 97:Issue 4(2017)
- Issue Display:
- Volume 97, Issue 4 (2017)
- Year:
- 2017
- Volume:
- 97
- Issue:
- 4
- Issue Sort Value:
- 2017-0097-0004-0000
- Page Start:
- 338
- Page End:
- 342
- Publication Date:
- 2017-12
- Subjects:
- Dental handpieces -- Steam sterilization -- Non-vacuum sterilizers -- Vacuum sterilizers -- Dental infection control -- Thermocouples
Cross infection -- Periodicals
Cross infection -- Prevention -- Periodicals
Nosocomial infections -- Periodicals
Nosocomial infections -- Prevention -- Periodicals
Cross Infection -- Periodicals
Cross Infection -- prevention & control -- Periodicals
Infection Control -- Periodicals
Electronic journals
614.44 - Journal URLs:
- http://www.harcourt-international.com/journals ↗
http://www.clinicalkey.com/dura/browse/journalIssue/01956701 ↗
http://www.sciencedirect.com/science/journal/01956701 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jhin.2017.07.033 ↗
- Languages:
- English
- ISSNs:
- 0195-6701
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5003.285000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5486.xml