Determination of background, signal‐to‐noise, and dynamic range of a flow cytometer: A novel practical method for instrument characterization and standardization. Issue 11 (27th September 2017)
- Record Type:
- Journal Article
- Title:
- Determination of background, signal‐to‐noise, and dynamic range of a flow cytometer: A novel practical method for instrument characterization and standardization. Issue 11 (27th September 2017)
- Main Title:
- Determination of background, signal‐to‐noise, and dynamic range of a flow cytometer: A novel practical method for instrument characterization and standardization
- Authors:
- Giesecke, Claudia
Feher, Kristen
von Volkmann, Konrad
Kirsch, Jenny
Radbruch, Andreas
Kaiser, Toralf - Abstract:
- Abstract: A well‐defined scale calibration in flow cytometry can improve many aspects of data acquisition such as cytometer setup, instrument comparison and sample comparison. The theory for scale calibration was proposed by Steen over two decades ago, but it has never been put into regular use due to the lack of a widely available precision light source. The introduction of such a light source, the quantiFlash TM, gave this possibility. Here, we describe how this light source can be used to characterize a cytometer's PMT performance. We, therefore, characterized the instrument's response over the entire PMT voltage range. As a consequence, we propose a practical method to characterize a cytometer's signal‐to‐noise ratio (SNR) and dynamic range (DNR). This allows the selection of a voltage/gain corresponding to a PMT's maximum efficiency and hence the lowest electronic noise, which can help with experiment design. We further introduced a decibel (dB) scale for the presentation of SNR and DNR values. SNR and DNR are stand‐alone values that allow the direct comparison of different instruments. Finally, with this method, it becomes clear that increased SNR comes at the expense of DNR and thus the limiting factor of modern cytometers is the DNR. © 2017 International Society for Advancement of Cytometry
- Is Part Of:
- Cytometry. Volume 91:Issue 11(2017)
- Journal:
- Cytometry
- Issue:
- Volume 91:Issue 11(2017)
- Issue Display:
- Volume 91, Issue 11 (2017)
- Year:
- 2017
- Volume:
- 91
- Issue:
- 11
- Issue Sort Value:
- 2017-0091-0011-0000
- Page Start:
- 1104
- Page End:
- 1114
- Publication Date:
- 2017-09-27
- Subjects:
- calibration -- characterization -- standardization -- LED light source -- background light -- signal‐to‐noise ratio -- dynamic range
Flow cytometry -- Periodicals
Imaging systems in biology -- Periodicals
Imaging systems in medicine -- Periodicals
Diagnostic imaging -- Periodicals
571.605 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1552-4930 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/cyto.a.23250 ↗
- Languages:
- English
- ISSNs:
- 1552-4922
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3506.855100
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5546.xml