Accuracy, precision and inter-rater reliability of micro-CT analysis of false starts on bones. A preliminary validation study. (November 2017)
- Record Type:
- Journal Article
- Title:
- Accuracy, precision and inter-rater reliability of micro-CT analysis of false starts on bones. A preliminary validation study. (November 2017)
- Main Title:
- Accuracy, precision and inter-rater reliability of micro-CT analysis of false starts on bones. A preliminary validation study
- Authors:
- Pelletti, Guido
Cecchetto, Giovanni
Viero, Alessia
Fais, Paolo
Weber, Michael
Miotto, Diego
Montisci, Massimo
Viel, Guido
Giraudo, Chiara - Abstract:
- Graphical abstract: Highlights: Micro-CT was tested for the analyses of false starts on bones. Three independent blind raters performed qualitative and quantitative analyses. A 100% morphologic agreement and a high level of accuracy (error < 0, 5%) were observed. Precision (CV% < 5) and inter-rater reliability (ICC > 0.995) indicated high reproducibility. Further validation studies are needed using a higher number of samples and tools. Abstract: Micro computed tomography (micro-CT) has already been proposed as a useful technique for the qualitative analysis of false starts (FS) produced on human bones, although the reliability and the error rate of this technique have not been tested yet, neither for qualitative nor for quantitative assessments. The aim of the present study was to test the morphological agreement, accuracy, precision and inter-rater reliability of micro-CT analysis of FS on bones. The morphological agreement was assessed through the degree of concordance among the 3 independent blind raters in the identification of the shape of 24 FS manually produced on bones by 3 different saws (8 FS for each saw). The accuracy was calculated through the percentage of error in the automatic and manual measurement of the diameter of a reference object. The precision was calculated as CV% of multiple measurements performed by 3 independent blind raters on the reference object and one bone sample acquired 20 times. The inter-rater reliability was assessed as intraclassGraphical abstract: Highlights: Micro-CT was tested for the analyses of false starts on bones. Three independent blind raters performed qualitative and quantitative analyses. A 100% morphologic agreement and a high level of accuracy (error < 0, 5%) were observed. Precision (CV% < 5) and inter-rater reliability (ICC > 0.995) indicated high reproducibility. Further validation studies are needed using a higher number of samples and tools. Abstract: Micro computed tomography (micro-CT) has already been proposed as a useful technique for the qualitative analysis of false starts (FS) produced on human bones, although the reliability and the error rate of this technique have not been tested yet, neither for qualitative nor for quantitative assessments. The aim of the present study was to test the morphological agreement, accuracy, precision and inter-rater reliability of micro-CT analysis of FS on bones. The morphological agreement was assessed through the degree of concordance among the 3 independent blind raters in the identification of the shape of 24 FS manually produced on bones by 3 different saws (8 FS for each saw). The accuracy was calculated through the percentage of error in the automatic and manual measurement of the diameter of a reference object. The precision was calculated as CV% of multiple measurements performed by 3 independent blind raters on the reference object and one bone sample acquired 20 times. The inter-rater reliability was assessed as intraclass correlation coefficient (ICC) among measurements performed by 3 independent blind raters, assessing 24 FS produced using 3 different saws. The results demonstrated that both qualitative and quantitative analysis were reproducible and robust. Micro-CT analysis showed a 100% morphological agreement, a high level of accuracy (percentage error < 0, 5%), precision (CV% < 5%) and inter-rater reliability (ICC > 0.995), when FS were analyzed by forensic pathologists and/or radiologists with adequate expertise. Obviously, further validation studies are needed, including a higher number of samples produced by a wider variety of saws and multiple operators. … (more)
- Is Part Of:
- Legal medicine. Volume 29(2017)
- Journal:
- Legal medicine
- Issue:
- Volume 29(2017)
- Issue Display:
- Volume 29, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 29
- Issue:
- 2017
- Issue Sort Value:
- 2017-0029-2017-0000
- Page Start:
- 38
- Page End:
- 43
- Publication Date:
- 2017-11
- Subjects:
- Forensic anthropology -- False starts -- Micro-CT -- Qualitative and quantitative assessment -- Validation study
Medical jurisprudence -- Periodicals
Forensic Medicine -- Periodicals
Médecine légale -- Périodiques
Medical jurisprudence
Periodicals
614.1 - Journal URLs:
- http://www.sciencedirect.com/science/journal/13446223 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.legalmed.2017.10.003 ↗
- Languages:
- English
- ISSNs:
- 1344-6223
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5181.329970
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5376.xml