Assessment of a photon‐counting detector for a dual‐energy C‐arm angiographic system. Issue 11 (22nd September 2017)
- Record Type:
- Journal Article
- Title:
- Assessment of a photon‐counting detector for a dual‐energy C‐arm angiographic system. Issue 11 (22nd September 2017)
- Main Title:
- Assessment of a photon‐counting detector for a dual‐energy C‐arm angiographic system
- Authors:
- Ahmad, Moiz
Fahrig, Rebecca
Pung, Leland
Spahn, Martin
Köster, Niko S.
Reitz, Silke
Moore, Teri
Choi, Jang‐Hwan
Hinshaw, Waldo
Xia, Yan
Müller, Kerstin - Abstract:
- Abstract : Purpose: This article presents the implementation and assessment of photon‐counting dual‐energy x‐ray detector technology for angiographic C‐arm systems in interventional radiology. Methods: A photon‐counting detector was successfully integrated into a clinical C‐arm CT system. Detector performance was assessed using image uniformity metrics in both 2D projections and 3D cone‐beam computed tomography (CBCT) images. Uniform exposure fields were acquired to analyze projection images and scans of a homogeneous cylinder phantom were taken to analyze 3D reconstructions. Image uniformity was assessed over a broad range of imaging parameters. Results: Detector calibration greatly improved image uniformity, reducing image variation from 8.8% to 0.5% in an ideal scenario, but image uniformity degraded when imaging parameters varied strongly from values set at calibration: the tube voltage, low‐high energy threshhold, and tube current had the greatest impact. Material discrimination and dynamic angiography capabilities were successfully demonstrated in separate phantom and in vivo experiments. Conclusion: The uniformity results identified major factors degrading image quality. The quantitative results will guide selection of calibration points to mitigate the loss of uniformity. The unique combination of dual‐energy and fluoroscopy imaging capabilities with a flat‐panel photon‐counting detector may enable new applications in interventional radiology.
- Is Part Of:
- Medical physics. Volume 44:Issue 11(2017)
- Journal:
- Medical physics
- Issue:
- Volume 44:Issue 11(2017)
- Issue Display:
- Volume 44, Issue 11 (2017)
- Year:
- 2017
- Volume:
- 44
- Issue:
- 11
- Issue Sort Value:
- 2017-0044-0011-0000
- Page Start:
- 5938
- Page End:
- 5948
- Publication Date:
- 2017-09-22
- Subjects:
- C‐arm -- computed tomography -- dual‐energy -- photon‐counting -- x‐ray detectors
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.12517 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5361.xml