Evaluation of the performance of elastomeric pumps in practice: are we under-delivering on chemotherapy treatments?. (2nd December 2017)
- Record Type:
- Journal Article
- Title:
- Evaluation of the performance of elastomeric pumps in practice: are we under-delivering on chemotherapy treatments?. (2nd December 2017)
- Main Title:
- Evaluation of the performance of elastomeric pumps in practice: are we under-delivering on chemotherapy treatments?
- Authors:
- Salman, Dahlia
Biliune, Jurga
Kayyali, Reem
Ashton, Jane
Brown, Peter
McCarthy, Tim
Vikman, Elin
Barton, Stephen
Swinden, Julian
Nabhani-Gebara, Shereen - Abstract:
- Abstract: Background and aims: Elastomeric pumps are widely used to facilitate ambulatory chemotherapy, and studies have shown that they are safe and well received by patients. Despite these advantages, their end of infusion time can fluctuate significantly. The aim of this research was to observe the performance of these pumps in real practice and to evaluate patients' satisfaction. Methods: This was a two-phase study conducted at three cancer units over 6 months. Phase-1 was an observational study recording the status of pumps at the scheduled disconnection time and noting remaining volume of infusion. Phase-2 was a survey of patients and their perception/satisfaction. Ethical approval was granted. Results: A total of 92 cases were observed covering 50 cases disconnected at hospital and 42 disconnected at home. The infusion in 40% of hospital disconnection cases was slow, with patients arriving at hospital with unfinished pumps; 58% of these had an estimated remaining volume which exceeded 10 mL with 35% exceeded 20 mL. In 73% of these cases, and regardless of the remaining volume, the patient was disconnected and the pump was discarded. Conclusions: The performance of pumps varied, which affected nurse workload and patients' waiting-times. A smart system is an option to monitor the performance of pumps and to predict their accuracy.
- Is Part Of:
- Current medical research and opinion. Volume 33:Number 12(2017)
- Journal:
- Current medical research and opinion
- Issue:
- Volume 33:Number 12(2017)
- Issue Display:
- Volume 33, Issue 12 (2017)
- Year:
- 2017
- Volume:
- 33
- Issue:
- 12
- Issue Sort Value:
- 2017-0033-0012-0000
- Page Start:
- 2153
- Page End:
- 2159
- Publication Date:
- 2017-12-02
- Subjects:
- Ambulatory chemotherapy -- ambulatory infusion pump -- elastomeric pumps -- safety -- colorectal cancer -- fluorouracil -- 5-FU -- FOLFIRI -- FOLFOX
Clinical medicine -- Periodicals
Therapeutics -- Periodicals
615.5 - Journal URLs:
- http://informahealthcare.com ↗
- DOI:
- 10.1080/03007995.2017.1374936 ↗
- Languages:
- English
- ISSNs:
- 0300-7995
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3500.301000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5338.xml