Cardiorespiratory Mechanical Simulator for In Vitro Testing of Impedance Minute Ventilation Sensors in Cardiac Pacemakers. Issue 2 (March 2016)
- Record Type:
- Journal Article
- Title:
- Cardiorespiratory Mechanical Simulator for In Vitro Testing of Impedance Minute Ventilation Sensors in Cardiac Pacemakers. Issue 2 (March 2016)
- Main Title:
- Cardiorespiratory Mechanical Simulator for In Vitro Testing of Impedance Minute Ventilation Sensors in Cardiac Pacemakers
- Authors:
- Marcelli, Emanuela
Cercenelli, Laura - Abstract:
- Abstract : We developed a cardiorespiratory mechanical simulator (CRMS), a system able to reproduce both the cardiac and respiratory movements, intended to be used for in vitro testing of impedance minute ventilation (iMV) sensors in cardiac pacemakers. The simulator consists of two actuators anchored to a human thorax model and a software interface to control the actuators and to acquire/process impedance signals. The actuators can be driven separately or simultaneously to reproduce the cardiac longitudinal shortening at a programmable heart rate and the diaphragm displacement at a programmable respiratory rate (RR). A standard bipolar pacing lead moving with the actuators and a pacemaker case fixed to the thorax model have been used to measure impedance ( Z ) variations during the simulated cardiorespiratory movements. The software is able to discriminate the low-frequency component because of respiration ( Z R ) from the high-frequency ripple because of cardiac effect ( Z C ). Impedance minute ventilation is continuously calculated from Z R and RR. From preliminary tests, the CRMS proved to be a reliable simulator for in vitro evaluation of iMV sensors. Respiration impedance recordings collected during cardiorespiratory movements reproduced by the CRMS were comparable in morphology and amplitude with in vivo assessments of transthoracic impedance variations.
- Is Part Of:
- ASAIO journal. Volume 62:Issue 2(2016)
- Journal:
- ASAIO journal
- Issue:
- Volume 62:Issue 2(2016)
- Issue Display:
- Volume 62, Issue 2 (2016)
- Year:
- 2016
- Volume:
- 62
- Issue:
- 2
- Issue Sort Value:
- 2016-0062-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-03
- Subjects:
- minute ventilation -- mechanical simulator -- transthoracic impedance -- cardiorespiratory dynamics
Artificial organs -- Periodicals
617 - Journal URLs:
- http://journals.lww.com/asaiojournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAT.0000000000000302 ↗
- Languages:
- English
- ISSNs:
- 1058-2916
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1738.840500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5243.xml