Repeatability, Reproducibility, and Age Dependency of Dynamic Scheimpflug-Based Pneumotonometer and Its Correlation With a Dynamic Bidirectional Pneumotonometry Device. Issue 1 (January 2015)
- Record Type:
- Journal Article
- Title:
- Repeatability, Reproducibility, and Age Dependency of Dynamic Scheimpflug-Based Pneumotonometer and Its Correlation With a Dynamic Bidirectional Pneumotonometry Device. Issue 1 (January 2015)
- Main Title:
- Repeatability, Reproducibility, and Age Dependency of Dynamic Scheimpflug-Based Pneumotonometer and Its Correlation With a Dynamic Bidirectional Pneumotonometry Device
- Authors:
- Bak-Nielsen, Sashia
Pedersen, Iben Bach
Ivarsen, Anders
Hjortdal, Jesper - Abstract:
- Abstract : Purpose: To assess the repeatability, reproducibility, and age dependency of dynamic Scheimpflug-based pneumotonometry (Corvis)–generated parameters and their correlation with dynamic bidirectional pneumotonometry device [Ocular Response Analyzer (ORA)] measurements [corneal hysteresis (CH) and corneal resistance factor (CRF)]. Methods: Eyes of 29 healthy participants between 20 and 30 years old were measured using Corvis and ORA. Measurements were performed twice in a row and repeated after 1 week. Repeatability and reproducibility were examined by calculating intraday and interday coefficients of variation (CVs). Corvis measurements were compared with measurements from 19 participants older than 65 years. Results: The most repeatable and reproducible Corvis-generated parameters, with intraday and interday CVs < 10% were as follows: intraocular pressure (IOP), pachymetry, time to first and second applanation (A1 time, A2 time), radius and amplitude at highest concavity (HC radius and HC deformation amplitude, respectively), time to highest concavity (HC time), deflection length and amplitude at highest concavity [HC deflection length and HC deflection amplitude (millimeter), respectively], and time to highest deflection amplitude [HC deflection amplitude (millisecond)]. However, 7 of 19 analyzed parameters showed CVs > 10%. HC-related parameters showed significant age dependency. The correlation between Corvis and ORA was weak (A2 time and CH, r = 0.36, A1 timeAbstract : Purpose: To assess the repeatability, reproducibility, and age dependency of dynamic Scheimpflug-based pneumotonometry (Corvis)–generated parameters and their correlation with dynamic bidirectional pneumotonometry device [Ocular Response Analyzer (ORA)] measurements [corneal hysteresis (CH) and corneal resistance factor (CRF)]. Methods: Eyes of 29 healthy participants between 20 and 30 years old were measured using Corvis and ORA. Measurements were performed twice in a row and repeated after 1 week. Repeatability and reproducibility were examined by calculating intraday and interday coefficients of variation (CVs). Corvis measurements were compared with measurements from 19 participants older than 65 years. Results: The most repeatable and reproducible Corvis-generated parameters, with intraday and interday CVs < 10% were as follows: intraocular pressure (IOP), pachymetry, time to first and second applanation (A1 time, A2 time), radius and amplitude at highest concavity (HC radius and HC deformation amplitude, respectively), time to highest concavity (HC time), deflection length and amplitude at highest concavity [HC deflection length and HC deflection amplitude (millimeter), respectively], and time to highest deflection amplitude [HC deflection amplitude (millisecond)]. However, 7 of 19 analyzed parameters showed CVs > 10%. HC-related parameters showed significant age dependency. The correlation between Corvis and ORA was weak (A2 time and CH, r = 0.36, A1 time and CRF, r = 0.41). Corvis-derived IOP was on average 2.2 ± 2.3 mm Hg higher than cornea-compensated IOP, with large differences in patients with high-average IOP. Conclusions: The considerable variation in repeatability and reproducibility should be taken into account when using the Corvis. Corvis parameters showed poor correlation with CH and CRF obtained by ORA measurements. … (more)
- Is Part Of:
- Cornea. Volume 34:Issue 1(2015)
- Journal:
- Cornea
- Issue:
- Volume 34:Issue 1(2015)
- Issue Display:
- Volume 34, Issue 1 (2015)
- Year:
- 2015
- Volume:
- 34
- Issue:
- 1
- Issue Sort Value:
- 2015-0034-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-01
- Subjects:
- cornea biomechanics -- Scheimflug-imaging technic -- Corvis ST -- ocular response analyzer -- reproducibility -- age dependency
Cornea -- Periodicals
Cornea -- Periodicals
Cornée -- Périodiques
617.719 - Journal URLs:
- http://journals.lww.com/corneajrnl/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/ICO.0000000000000293 ↗
- Languages:
- English
- ISSNs:
- 0277-3740
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3470.927500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5251.xml